49
Example 1:
Count parts into a container
Menu settings: factory settings
Step
Press key
Display/Printout
1. Turn on balance,
[ON/OFF]
if power is off.
2. Place empty container
[START]
0.0
g
on balance and start counting.
+
36.6 SAMPLE
3. Select reference sample
Brief display of:
quantity; for example,
turn selector to “20”.
ref 20
4. Place reference sample
quantity on balance
+
23.1
g
(20 parts in this case).
5. Save reference sample quantity.
[OK]
+
20
pcs
The balance now calculates the
+
23.1
+
1.1550
average piece weight.
6. If necessary, update the reference
sample using the optimizing function;
+
32
pcs
in this case, add 2 to 20 additional
+
36.6
opt.
parts to the weighing pan.
7. Now place unknown number of parts
+
260
pcs
on pan (in this example, 260).
+
297,4
+
1.1438
8. Print piece count.
nRef
+
32
If you have selected
[PRINT]
wRef
+
1.1438 g
“nRef” and “wRef” in the menu,
Qnt
+
260 pcs
these lines will be printed:
G
+
297.4 g
9. Unload balance
0
pcs
10. Repeat counting starting
from step 7, if desired.
PK Elektronik Vertriebs GmbH, E-Mail: [email protected], Internet: www.pkelektronik.com