7.2.2 P1.2 ECAL - External adjustment (AES models)
Carry out adjustment with the help of the recommended adjustment weight (see
chapter 1 “Technical Specifications”).
Observe stable environmental conditions. A warming up time (see chapter 1) is
required for stabilization.
•
Return to weighing mode, see chap. 8.2.
•
This function is locked in weighing scales with type approval certificate.
Ensure that there are no objects on the
weighing pan. Press
-key,
CAL
will be shown.
Wait until the weighed value for the
required adjustment weight appears.
Place the required calibration weight
carefully in the centre of the weighing
plate and press the
key.
The display is showing
CAL
and
adjustment is starting.
Adjustment will be complete when the
display is showing
UNLOAD
; remove
the adjustment weight. After successful
adjustment the balance automatically
returns to weighing mode.