5  Troubleshooting 

 
 

Interruption 

Possible cause 

==

OVERLOAD

== 

 

  Either the scale is overloaded or there is a mechanical 

defect 

= ============ 

 

  Underload: Weighing range has been fallen short of 

  Mechanical defect or a defect in the main PCB 

           

<<

            . 

 

  In counting and % mode: part too light 

           

bat

            . 

 

  The rechargeable battery is empty and should be 

loaded 

Weight display is not illuminated. 

  The balance is not switched on. 

 

  The mains supply connection has been interrupted 

(mains cable not plugged in/faulty). 

 

  Power supply interrupted. . 

The weight display changes continually   

 Draught/air 

movement 

 

  Table/floor vibrations  

 

  The weighing plate is in contact with foreign matter. 

  Electromagnetic fields / static charging (choose different 

location/switch off interfering device if possible) 

The weighing result is obviously incorrect 

  The balance display is not set to zero  

 

  Adjustment is no longer correct. 

 

  Great fluctuations in temperature. 

  Electromagnetic fields / static charging (choose different 

location/switch off interfering device if possible) 

 

 
Switch the balance off if other error messages should appear and then switch on again. 
Contact the manufacturer if the error message does not disappear.  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for 572xxNM

Page 1: ...D 72336 Balingen E Mail info kern sohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com Service manual Precision balances KERN 572 573xxNM Version 5 1 KBxxN M DS GB FCBxxB F...

Page 2: ...5 Troubleshooting 6 6 Adjustment 7 6 1 Adjustment of non verification balances 7 6 2 Adjustment of verification balances KERN 573 NM KB NM FKB M 7 7 Factory Menu 8 7 1 How to Enter the Factory Menu 8...

Page 3: ...l After successful repair the balance will have to be reverified before it can be used again in a statutorily regulated field Das Ger t darf nur von geschultem oder beruflich ausgebildetem Fachpersona...

Page 4: ...d Copying any part of this manual is prohibited without our permission 3 Keyboard Description Keyboard ON OFF Print weighing result in MODE No descending in and counting mode Form reference in MODE Ye...

Page 5: ...4 Mode Menu IMPORTANT The modified settings as well as the adjustment must be stored when switching off via the ON OFF button 572 SH e 0951 5...

Page 6: ...Power supply interrupted The weight display changes continually Draught air movement Table floor vibrations The weighing plate is in contact with foreign matter Electromagnetic fields static charging...

Page 7: ...step is located at the bottom of the housing between the two turning feet Switch to the right Adjusting function is released This setting has to be selected before starting the adjusting Executing the...

Page 8: ...once to return to weighing mode Please notice the LCD segments are intense visible if the stored value is low If the stored value is high the LCD segments are low visible 7 3 Factory Adjustment 1 Turn...

Page 9: ...iguration software 7 5 Manual Linearity Adjustment 1 Turn ON the balance After warming up time and adjustment check balance readings in all measuring points In case of differences make a note 2 Press...

Page 10: ...oard overlay JTAG 2 3 4 5 1 Connection of LCD display Load cell IN EX IN EX Shield OUT SIG OUT SIG Connection of LCD display Series resistor for backlight min 50 ohm Top Side If required set solder br...

Page 11: ...8 2 Layout of Main PCB 572 SH e 0951 11...

Reviews: