5 Troubleshooting
Interruption
Possible cause
==
OVERLOAD
==
•
Either the scale is overloaded or there is a mechanical
defect
= ============
•
Underload: Weighing range has been fallen short of
•
Mechanical defect or a defect in the main PCB
<<
.
•
In counting and % mode: part too light
bat
.
•
The rechargeable battery is empty and should be
loaded
Weight display is not illuminated.
•
The balance is not switched on.
•
The mains supply connection has been interrupted
(mains cable not plugged in/faulty).
•
Power supply interrupted. .
The weight display changes continually
•
Draught/air
movement
•
Table/floor vibrations
•
The weighing plate is in contact with foreign matter.
•
Electromagnetic fields / static charging (choose different
location/switch off interfering device if possible)
The weighing result is obviously incorrect
•
The balance display is not set to zero
•
Adjustment is no longer correct.
•
Great fluctuations in temperature.
•
Electromagnetic fields / static charging (choose different
location/switch off interfering device if possible)
Switch the balance off if other error messages should appear and then switch on again.
Contact the manufacturer if the error message does not disappear.