7.4 Automatic Linearity Adjustment 

 

1.  Turn ON the balance. 
2.  Press and hold the TARE key approx. 10 seconds until [

Factory Reset?

appears. 

3.  Press MODE key until [

Linear auto ?

] appears. 

4. Press 

 

key to start the automatic linearity adjustment. The display appears 

the first linearity point, e.g. [

L1  2000.0g  ?

]. 

5. Confirm 

with 

 

key or press the MODE key to select another linearity point, 

which is outside the permissible tolerances, e.g. [

L2  5000.0g  ?

] and confirm with 

 

key. The display appears [

Place Lin Weight

]. 

6.  Place the required weight on the centre of the weighing pan and wait until you will 

get a stable weighing result.  

7.  Remove the weight. Now balance is in weighing mode. 
8. 

Switch off the balance by pressing the ON/OFF-key.

 

9. 

Now plug off the power supply and plug in again.

 

 
Please notice the linearity points were defined and configured in our production line. 
Therefore it is not possible to change them afterwards without special configuration 
software. 
 

7.5 Manual Linearity Adjustment 

 

1.  Turn ON the balance. After warming-up time and adjustment check balance 

readings in all measuring points. In case of differences make a note. 

2.  Press and hold the TARE key approx. 10 seconds until [

Factory Reset?

appears. 

3.  Press MODE key until [

Linear man ?

] appears. 

4. Press 

 

key to start the manual linearity adjustment. The display appears the 

first linearity point, e.g. [

L  2000.0g  +00d

]. 

5.  With the arrow keys 

 

and

 

 

you can change the stored value. 

6.  Press the MODE key to select another linearity point, which is outside the 

permissible tolerances, e.g. [

L  5000.0g  -07d

] and inscribe the correction value 

like described above in point (5.) and so on. 

7.  To return to normal weighing mode press TARE key once. 
8. 

Switch off the balance by pressing the ON/OFF-key.

 

9. 

Now plug off the power supply and plug in again.

 

 

Information

If there is a positive difference (e.g. 2000.6 g = + 6 digit) you should subtract this difference 
to the stored value (e.g. [L  2000.0  -06d]). 

 

If there is a negative difference (e.g. 1999.4 g = - 6 digit) you should add this difference to 
the stored value (e.g. [L 2000.0  +06d]). 

572-SH-e-0951 

                                   

        

9  

 

 

Summary of Contents for 572xxNM

Page 1: ...D 72336 Balingen E Mail info kern sohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com Service manual Precision balances KERN 572 573xxNM Version 5 1 KBxxN M DS GB FCBxxB F...

Page 2: ...5 Troubleshooting 6 6 Adjustment 7 6 1 Adjustment of non verification balances 7 6 2 Adjustment of verification balances KERN 573 NM KB NM FKB M 7 7 Factory Menu 8 7 1 How to Enter the Factory Menu 8...

Page 3: ...l After successful repair the balance will have to be reverified before it can be used again in a statutorily regulated field Das Ger t darf nur von geschultem oder beruflich ausgebildetem Fachpersona...

Page 4: ...d Copying any part of this manual is prohibited without our permission 3 Keyboard Description Keyboard ON OFF Print weighing result in MODE No descending in and counting mode Form reference in MODE Ye...

Page 5: ...4 Mode Menu IMPORTANT The modified settings as well as the adjustment must be stored when switching off via the ON OFF button 572 SH e 0951 5...

Page 6: ...Power supply interrupted The weight display changes continually Draught air movement Table floor vibrations The weighing plate is in contact with foreign matter Electromagnetic fields static charging...

Page 7: ...step is located at the bottom of the housing between the two turning feet Switch to the right Adjusting function is released This setting has to be selected before starting the adjusting Executing the...

Page 8: ...once to return to weighing mode Please notice the LCD segments are intense visible if the stored value is low If the stored value is high the LCD segments are low visible 7 3 Factory Adjustment 1 Turn...

Page 9: ...iguration software 7 5 Manual Linearity Adjustment 1 Turn ON the balance After warming up time and adjustment check balance readings in all measuring points In case of differences make a note 2 Press...

Page 10: ...oard overlay JTAG 2 3 4 5 1 Connection of LCD display Load cell IN EX IN EX Shield OUT SIG OUT SIG Connection of LCD display Series resistor for backlight min 50 ohm Top Side If required set solder br...

Page 11: ...8 2 Layout of Main PCB 572 SH e 0951 11...

Reviews: