572/573/KB/DS/FKB/FCB/KBJ-BA-e-1370
21
KERN
FKB
8K0.1M
FKB
65K1M
Accuracy class
II
II
Readability (d)
0.1 g
1 g
Verification value (e)
1.0 g
10 g
Weighing range (max)
8 000 g
65 000 g
Minimum load (Min)
5.0 g
50 g
Taring range (subtractive)
8 000 g
65 000 g
Reproducibility
0.1 g
1 g
Linearity
± 0.3 g
± 3 g
Smallest piece weight
0.1 g
1 g
Adjustment points
2/5/7/8 kg
20/30/50/60 kg
Recommended adjusting
weight F1 (not supplied)
5 kg + 2 kg
50 kg
Humidity of air
max. 80% rel. (non-condensing)
Stabilization time (typical)
3 sec.
Permitted environmental
temperature
+10 °C ... + 30 °C
+10 °C ... + 30 °C
Warm-up time
2 hours
2 hours
Housing (B x D x H) mm
350 x 390 x 120
Vibration filter
yes
Weighing plate stainless
steel mm
340 x 240
340 x 240
Units
see menu
see menu
Weight kg (net)
6,5
6.5
Data interface
yes (RS232)