2-12

MAT 6V/15V/25V/36V/ 022300

d.

The MAT previously set up now starts to measure and send the proper time delay
needed for the measurement.

e.

The  Power  Module  Controller  will send  the   time  delay  value back  to  the  Host  Com-
puter when it has been  properly  Talk  Addressed  (see  specific  IEEE  488  interface
requirements):
00(cr)(lf)

1.

The Host Computer then sends:

FTH  CURR(cr)(lf)

a.

The MAT then sends the measured current in scientific notation to the Power Module
Controller (for example 9.9998 amps) which

b.

The Power Module Controller  in turn sends

 9.9998E0(cr)(lf)

3.

Check that the front panel meter reading and the computer are identical within their mea-
surement tolerance. The meter selection switch must be set to "Current" for a reading.

2.8.3

EXAMPLE 3: TO RESET THE POWER MODULE

4.

 To reset the Power Module at address 3 the Host Computer sends:

RST  DCS  :CH3(cr)(lf)

5.

To verify the RST command was implemented, the Host Computer sends:

STA(cr)(lf)

The Power Module Controller sends back (if no errors occur):

(sp)(cr)(lf)

6.

 This concludes the preliminary check-out  of the MAT Power Module.

Summary of Contents for MAT FULL RACK 15V

Page 1: ...t revision for your par ticular model and serial number 3 The contents of this manual are protected by copyright Reproduction of any part can be made only with the specific written permission of Kepco...

Page 2: ......

Page 3: ...USA Importer s Name and Address Type of Equipment Component Power Supply Model No PRODUCT MODEL NUMBER Year of Manufacture I the undersigned declare that the product specified above when used in conj...

Page 4: ...wire grounded mains connection See wiring section of this manual for complete electrical wiring instructions EN61010 1 2001 Cl 6 10 1 4 This power supply has secondary output circuits that are conside...

Page 5: ...connection on the product s rear panel To reconfigure the product input for other nominal mains voltages as listed herein the product must be modified by a trained service technician The PROTECTIVE E...

Page 6: ...2 9 2 8 1 Example 1 Full Scale Voltage Check 2 10 2 8 2 Example 2 Full Scale Current Check 2 11 2 8 3 Example 3 To Reset the Power Module 2 12 SECTION 3 OPERATION 3 1 General 3 1 3 2 Introduction To...

Page 7: ...dance With Frequency For A Voltage Source And A Current Source 2 7 2 5 Load Connections Local Sensing 2 8 2 6 Load Connections Remote Sensing 2 8 3 1 Top Cover Access Points 3 1 3 2 MAT Power Module C...

Page 8: ...9 2 1 Internal Controls Accessible Through Top Cover 2 1 2 2 Front Control MAT Power Module 2 2 2 3 Rear Terminations MAT Power Module 2 3 2 4 DC Output Connector Pin Functions 2 3 2 5 Input And Outpu...

Page 9: ...oard control of the MAT via the IEEE 1118 bus b Controller Model TMA 4882 27 is free standing and allows host computers designed for RS232 or IEEE 488 bus communication to control the MAT via the IEEE...

Page 10: ...1 2 MAT FULL RACK 022300 FIGURE 1 1 REMOTELY CONTROLLED POWER SUPPLY CONFIGURATIONS USING KEPCO PRODUCTS 3040800...

Page 11: ...ater Time 8 Hour Drift Voltage 0 01 Eo max Current 0 02 Io max Temperature Coefficient C Voltage 0 01 Eo max Current 0 02 Io max Ripple Noise Voltage rms 0 001 Eo max or 0 3 mV whichever is greater p...

Page 12: ...ied and proper gage wires not supplied Load Connections 5 wire connection via user configurable mating connector supplied and proper gage wires not supplied DC Power Output 2 wires Output Sense Termin...

Page 13: ...0 3 6 360 MAT 150 2 4 0 150 0 2 4 360 720W MODULES 2 3 RACK DC OUTPUT RATINGS MAT 6 64 0 6 0 64 384 MAT 15 40 0 15 0 40 600 MAT25 28 0 25 0 28 700 MAT 36 20 0 36 0 20 720 MAT 55 14 0 55 0 14 770 MAT 7...

Page 14: ...1 6 MAT FULL RACK 022300 FIGURE 1 2 MAT POWER MODULE MECHANICAL OUTLINE DRAWING SHEET 1 OF 2...

Page 15: ...MAT FULL RACK 022803 1 7 FIGURE 1 2 MAT POWER MODULE MECHANICAL OUTLINE DRAWING SHEET 2 OF 2...

Page 16: ...es 1 6 ACCESSORIES a Kepco full rack MAT Models are drilled for optional chassis slides Kepco P N 108 0430 modified Jonathan 370 QD Series two required see Figure 1 2 b One 2 meter long shielded twist...

Page 17: ...t specified by Kepco Inc may impair the pro tection provided by the power supply Observe all safety precautions noted throughout this man ual Table 1 3 lists symbols used on the power supply or in thi...

Page 18: ...adjust output current of power supply to initial value CONDITIONS current mode shunt at the output A1R20 EO ZERO NOTE 1 Used to adjust output voltage of power supply to zero CONDITIONS voltage mode n...

Page 19: ...a few seconds before trying again The circuit breaker is trip free design if overload exists or the protection is activated contacts cannot be held closed by actuator 2 VOLTAGE MODE indicator green LE...

Page 20: ...2 2 for pin designations 2 CONTROL BUS CONNECTORS Connects MAT power module to the IEEE 1118 control bus Two connectors in parallel allow daisy chain connection of up to to 27 MAT power modules to be...

Page 21: ...the barrier terminal block set to 115 Va c operation Remove the jump ers on the barrier strip connecting terminals 2 to 3 and terminals 4 to 5 3 Place a jumper between terminals 3 and 4 For 230 Va c t...

Page 22: ...e the limit specified 2 5 INSTALLATION The MAT Power Module is designed to be rack mounted in a standard 19 inch equipment rack refer to Figure 1 2 Mechanical Outline Drawing For installation into con...

Page 23: ...ned to ground with a single wire This method is dependent on the specific application The MAT Power Modules are designed with Power Isolation relays that comply with MATE Modular Automated Test Equipm...

Page 24: ...tage drops error sensing is discussed below and minimum inductance error sensing does not compen sate for this Similarly a model for a current stabilized Power Module includes a parallel capaci tor re...

Page 25: ...NS LOCAL SENSING 2 6 5 LOAD CONNECTION WITH REMOTE ERROR SENSING The MAT series of Power Modules can operate with sensing external to the Module Sensing should be with a twisted wire pair to reduce no...

Page 26: ...llowing check out uses a MAT Power Module Model 100 10 as an example to the Control Bus Connect the Host Computer and the Power Module Controller to the IEEE 488 GPIB Apply the ac power to the units i...

Page 27: ...cr lf The Power Module Controller sends back if no errors occur sp cr lf 5 Proceed per PAR 2 8 1 through 2 8 3 substituting full scale voltage current values appropi ate to the model being checked 2 8...

Page 28: ...entical within the measurement tolerance The meter selection sewitch must be set to voltage for a reading 2 8 2 EXAMPLE 2 FULL SCALE CURRENT CHECK Program the Power Module MAT 100 10 to full scale cur...

Page 29: ...to the Power Module Controller for example 9 9998 amps which b The Power Module Controller in turn sends 9 9998E0 cr lf 3 Check that the front panel meter reading and the computer are identical withi...

Page 30: ...r Modules communicates remotely with a controller via the Control Bus Each MAT Power Module has a specific address on the bus The address from 1 to 31 is selected by the switches on the top of the uni...

Page 31: ...0 3 0 0 0 1 1 4 0 0 1 0 0 5 0 0 1 0 1 6 0 0 1 1 0 7 0 0 1 1 1 8 0 1 0 0 0 9 0 1 0 0 1 10 0 1 0 1 0 11 0 1 0 1 1 12 0 1 1 0 0 13 0 1 1 0 1 14 0 1 1 1 0 15 0 1 1 1 1 16 1 0 0 0 0 17 1 0 0 0 1 18 1 0 0 1...

Page 32: ...le Controller as an interface to the Host Computer for reporting catastrophic failures within the Power Modules If such an error is reported back to the Power Module Controller the Host Com puter then...

Page 33: ...his command see the applicable Power Module Controller Manual The required time delay before a valid status of the Power Module Controller MAT system is as follows After an Open Close Change of Polari...

Page 34: ...bus using either SCPI Standard Commands for Programmable Instruments or CIIL Control Interface Intermediate Language commands SCPI and CIIL provide a common language conforming to IEEE488 2 for instr...

Page 35: ...different See PAR 3 2 to establish the MAT power supply Control Bus BIT BUS address FIGURE 3 3 TREE DIAGRAM OF SCPI COMMANDS USED WITH MAT POWER SUPPLIES 3 6 2 1 SCPI PROGRAM EXAMPLE Figure 3 4 is an...

Page 36: ...DEV6 defined by IBCONF bd ibfind GPIB0 Open GPIB card ibsic bd Send Interface Clear ibsre bd 1 Set remote line true strcpy dat_str VOLT 5 CURR 1 Define a set command strcat dat_str r n Append delimite...

Page 37: ...efer to PAR 3 4 The examples are for a MAT 150 7 model with Control Bus address 9 The ASCII characters for space carriage return and line feed are represented as sp cr and lf respectively The MAT Powe...

Page 38: ...the current at the output the Host Computer sends FNC DCS CURR CH9 cr lf The MAT at address 9 is now set to take a reading 6 To take a reading the the Host Computer sends INX CURR cr lf a The MAT pre...

Page 39: ...nds FNC DCS CH9 SET VOLT 120 SET CURL 2 cr lf The MAT at address 9 is then programmed to 120 volts with a current limit of 2 amps 2 To verify the previous command was implemented the Host Computer sen...

Page 40: ...To reset the Power Module at address 9 the Host Computer sends RST DCS CH9 cr lf The MAT then programs the output to zero and opens the Output Enable relay 9 To verify the previous command was impleme...

Page 41: ...iously set up now starts to measure and sends the proper time delay needed for the measurement See Par 3 4 b The Power Module Controller will send the time delay back to the Host Computer when it has...

Page 42: ...mode Description The first operand contains the three 3 letter mnemonic pertaining to the device on the control bus in this case DCS Direct Current Source If a reading is being set up the modifier VOL...

Page 43: ...or the command may be repeated until a zero value is returned but the preceding FTH command must also be repeated Example INX VOLT Power supply initiates voltage reading FTH VOLT Power supply sends v...

Page 44: ...e MAT power supply allows 12 bits or 024 of maximum rated voltage or current It can be directly specified in ASCII integer decimal or in scientific notation There may be two 2 set commands separated b...

Page 45: ...a power supply to its power on state The output voltage and current are programmed to zero and the output relay of MR and MGR models is opened Example RST DCS CH13 The power supply at node address 13...

Page 46: ...e A shutdown occurred due to thermal causes F07 DCSnn DEV Overload The voltage or current limit point was exceeded F07 DCSnn DEV Voltage Fault The output voltage is not within limits voltage mode F07...