key erases data entered). Press ENTER to accept new
setting, or CLEAR to exit without changing setting.
NOTE: Blinking equal sign is indicated by in this manual.
4.2.3. DISPLAY (LCD).
The LCD is a 2-line display with a
capacity of 16 characters per line. The information is gen-
erally arranged as follows (information that does not follow
this format is self-explanatory): At top left:
Loc/Rem
indi-
cates Local or Remote status. At top right:
CV/CC
indi-
cates Constant Voltage or Constant Current mode. At
bottom left:
n.nnnV
indicates output voltage if in com-
mand entry or
(parameter
)
(e.g.
OVset
if
OV SET
key
was pressed) if in data entry. Bottom middle shows
(blinking “
:
”) for command entry or (blinking “
=
”) for data
entry. At bottom right:
n.nnnA
indicates output current in
command entry, or, in data entry,
n.nnn
indicates the
present value of the parameter indicated at the bottom left
(the present value is replaced by data entered).
4.2.4. KEYPAD FUNCTIONS.
(Refer to Figure 1.) Six
keys have dual functions, depending on whether the
power supply is in command entry status (waiting for a
command to be entered), or data entry status (waiting for
a number to be entered). Command entry status is
indicted on the LCD by a blinking colon and data entry
status is indicated by a blinking equal sign .
NOTE: Keys with dual functions are labeled with both a
command and a number. The command label is
referred to when the unit is in command entry
status; the number is referred to when the unit is
in data entry status.
OUTPUT ON/OFF
-
Command Entry active:
If bottom line of LCD reads
Output OFF
, press to
enable the output. If output is on (voltage and current
measurements displayed on bottom line of LCD), press to
disable the output.
V SET
-
Command Entry active:
Press to set output voltage. After
V SET
is pressed,
previous setting is displayed. Data entry required to enter
new value of output voltage; press
ENTER
to accept dis-
played value.
I
SET
-
Command Entry active:
Press to set output current. After
I
SET
is pressed,
previous setting is displayed. Data entry required to enter
new value of output current; press
ENTER
to accept dis-
played value.
LOCAL
-
Command Entry active:
If the power supply is in remote mode, keypad is dis-
abled except for
LOCAL
key. Press to enable keypad. If
LCD reads
KEYPAD LOCKED
, The
LOCAL
key is disabled
and can only be unlocked by remote operation.
RESET
-
Command Entry active:
Press to restore the power on default values: CV
mode, output voltage = 0, output current = minimum (1-
2% of
I
O
max), output enabled, overvoltage and overcur-
rent values. Also resets overvoltage or overcurrent condi-
tion.
MENU
-
Command Entry active:
Press to enter Menu commands: press repeatedly to
scroll through Menu functions: (1) set LCD contrast, (2)
GPIB address, (3) DCL Control, (4) Speaker on/off, (5)
Calibration password, (6) previous calibration values, (7)
factory calibration values, (8) view firmware serial number,
(9) set maximum voltage, (10) set maximum current, (11)
protection delay. Press
ENTER
or
RESET
to exit Menu.
OV SET / 7
Command Entry active:
Press to set overvoltage
protection value. Data entry required to enter the overvolt-
age protection value; press
ENTER
to accept displayed
value.
Data Entry active:
Press to enter number 7.
OC SET / 8
Command Entry active:
Press to set overcurrent
protection value. Data entry required to enter new over-
current protection value; press
ENTER
to accept dis-
played value.
Data Entry active:
Press to enter number 8.
CALIB / 9
Command Entry active:
Press to enter Calibration
status. Requires password entry; instructions appear on
LCD.
Data Entry active:
Press to enter number 9.
STORE
-
Command Entry active:
Press to store present values of output voltage and
current and overvoltage and overcurrent protection. Data
entry required to select memory location where values are
to be stored
EDIT PROG
-
Command Entry active:
Press to select the starting memory location to be
edited. Then use
or
to view or modify the parame-
ters of a specific memory location or to create a new pro-
gram.
STEP / 4
Command Entry active:
Press to select starting
address of program to be executed one step at a time.
Data
Entry active:
Press to enter number 4.