TK-378G
20
ADJUSTMENT
Fig. 1
BPF-WAVE
Fig. 2
Item
Condition
Measurement
Adjustment
Specifications/
Test equipment
Terminal
Parts
Method
Remark
M4 type
1. Sensitivity
f
L
:
Low
DVM
PC key
f
L
: [55]
Adjustment f
c
:
Center
f
C
: [100]
(PC MODE) f
H
:
High
f
H
: [160]
2. BPF
TEST CH : Low
Tracking generator
ANT
TC301, TC302, TC303
Adjust wave form to figure 1.(f
L
)
Tra-G setting -40dBm
spectrum analyzer
BPF
TEST CH : Center
Check wave form to figure 2.(f
C
)
TEST CH : High
Check wave form to figure 2.(f
H
)
3. Sensitivity
1) TEST CH : Center
SSG
ANT
Check
12dB SINAD or
Check
SSG OUT AF VTVM
more
Wide:-118dBm (0.28µV)
Oscilloscope
(MOD:1kHz / ±3kHz)
Narrow:-117dBm (0.316µV)
(MOD:1kHz / ±1.5kHz)
4. Squelch (Open) 1) TEST CH : Center
PC key
Adjust to the
Adjust
SSG OUT:
squelch
(PC MODE)
[Wide]
-123dBm (0.158µV)
threshold point
2) TEST CH : Center
[Narrow]
SSG OUT:
-122dBm (0.178µV)
5. Squelch(Tight) 1) TEST CH : Center
Adjust to the
Adjust
SSG OUT:-116dBm (0.35µV)
squelch
(PC MODE)
[Wide]
threshold point
[Narrow]
2) TEST CH : Center
SSG OUT:-115dBm (0.398µV)
6. Squelch
1) TEST CH : Center
Check
Squelch must
Check
SQ Level : [SQL5]
be opened.
(PC MODE)
SSG OUT:
-118dBm (0.28µV)
2) SSG OUT:OFF
Squelch must
be closed.