5. Product Main Features

 

6 Product Requirements and 

   Major Parametric Description

 

6.2 Product Main Technical Parametric description

a)Impulse frequency: 
TENS:(20-100)Hz, EMS:(1-15)Hz, FITNESS:(2-16)Hz.
b)Pulse width: 

TENS: 120μs, EMS: 200μs, FITNESS: 200μs.

c)Impulse waveform: square wave
d)The individual pulsing electric quantity when in the maximum output 

amplitude: >7μC.
e)The maximum output energy of individual pulse: ≤300mJ.

f)Influence of output end open circuit and short circuit: it is able to support 
the influence of output end open and short circuit, and its performance will 
not be allowed to weaken.
g)Adjustment of output amplitude: 0-16 levels.
h)The timer of therapeutic equipment: 15 minutes, 30 minutes, 45 minutes, 
±10% franchise.
i)Safety classification: internal electric source class and type BF equipment
j)Boundary dimension: 
Main unit: 122mm * 55mm * 15.7mm;
Electrode Pads: 3 kinds
EPAD-D01: 70×52mm        EPAD-D02: 50×50mm       EPAD-D03: 50×50mm
k)Service life: the shelf life of subject device is 2 years; the use life of the 
small Electrode Pad is 80 times and shelf life is 2 years

6.3 Product Environmental Requirement

a)Normal work environmental requirements:
Environment temperature: +5

-+40

Environment humidity: 15%-93%RH; 
Atmospheric environment conditions: 700hPa-1060hPa.
b)Storage environment requirements:
Environment temperature: -25

-+70

Environment humidity: 0-93%RH; Atmospheric 
Environment conditions: 700hPa-1060hPa.
c)Transport environment requirements:
Environment temperature: -25

-+70

Environment humidity: 0-93%RH; Atmospheric 
Environment conditions: 700hPa-1060hPa.
Note: It needs at least 30 minutes for device to return to normal operating 
temperature from the minimum/maximum storage temperature. 

The design is on the basis of ergonomics and is beautiful science, 

compact easy to carry;

The LF electric pulse come into being an efficient composite energy field;

Multi-function and multi-site remission, 

There are alternative functional modules, adjustable treatment time, and 

adjustable intensity;

There is LCD display;

04

05

6.1 Product Power Requirement

a) Power source: 3.7V/250mAh lithium battery
b) Device safety classes: class II type BF

Fig 1. Primary structure of main unit

Summary of Contents for KTR-2610

Page 1: ...us free of charge but the costs for parts and components and transport will be undertaken by the purchaser However man made damages such as water enters inside the product product drops on the ground...

Page 2: ...03 5 Product Main Features 04 6 Product Requirements and Major Parametric Description 04 7 Method of application 06 8 Safety Precautions 07 9 Product maintenance 10 10 Product Scrap Processing 11 11 A...

Page 3: ...arts have the haemorrhage after acute trauma or fractures disabling wound healing stage after the operation e Treatment is not sensitive to heat or electrical stimulation of the skin patients disabled...

Page 4: ...2 years the use life of the small Electrode Pad is 80 times and shelf life is 2 years 6 3 Product Environmental Requirement a Normal work environmental requirements Environment temperature 5 40 Envir...

Page 5: ...ase intensity 5 Mode switch Short press F key to switch three major modes of TENS EMS and FITNESS 6 The adjustment of timing Long press F key for 1 5 seconds to set up the stimulation time of 15 minut...

Page 6: ...l parts including 3 AAA LR03 battery Prevent sharp parts from damaging the product Do not use accessories not specified by the manufacturer Please do not use in the parts which is nearest heart head e...

Page 7: ...o move the plugs in emergency circumstances Check the equipment before each use to avoid the exposed wires caused by accidental damages or other reasons Dust may affect the performance of the unit ple...

Page 8: ...ooms near active AF surgical equipment electrophysiology laboratories armored rooms or areas where short wave therapy equipment is used Do not use the system around strong electric filed electromagnet...

Page 9: ...ommunications equipment including peripherals such as antenna cables and external antennas should by used no closer than 30cm to any part of the ultrasound system including cables specified by the man...

Page 10: ...17 Hz 2 2 0 3 0 3 0 2 0 3 0 3 28 28 9 1720 1845 1970 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE 1 3 4 25 Band UMTS Pulse modulation 217 Hz 2 28 2450 2400 2570 Bluetooth WLAN 802 1 1 b g n RFID 245...

Reviews: