User Manual
KS-PH-221-DELUXE |
Portable pH/ORP Meter
Deluxe
Kensington Scientific
22a Cavan Rd, Dry Creek
South Australia, 5094
[email protected]
www.kensingtonscientific.com.au
GENERAL OPTIONS
PARAMETER
DESCRIPTION
OPTIONS
DESCRIPTION
STA
Stable Criteria:
Sets the stability criteria for
measurement.
When the “LO” option is
enabled, measuring value will stabilize
quickly, but reading is less accurate.
When the “HI” option is enabled,
measuring value will stabilize slowly,
but guarantees high accuracy
LO (Default)
Low
HI
High
HOLD
Auto-Hold:
When the auto-hold function is
enabled, the meter will automatically
sense a stable end-point reading and
freeze it.
YES
Enable
NO (Default)
Disable
OFF
Auto-Power Off:
When the auto-off power is enabled, if
you do not press any key within a
specified time period, the meter will
automatically turn off.
10
10 minutes
20
20 minutes
30
30 minutes
NO (Default)
Disable
CALL
Calibration Due:
When calibration due reminder is
enabled, if you do not recalibrate meter
within a specified time period, the
meter will automatically show
indicator.
1…31
1 to 31 days
OFF (Default)
Disable
CLR
Clear stored data:
Clear all stored data
YES
Enable
NO (Default)
Disable
RST
Reset:
Reset function allows user to restore
the meter back to factory default
parameters. When this function is
enabled, all calibration values and
selected parameters will be lost or
reset.
YES
Enable