PREPARING THE pH BUFFER SOLUTIONS
The KS-PH-221-DELUXE comes with three pH buffer packets needed for calibration.
1. Begin by opening the packet for the 7.00 buffer. Pour distilled water into a 250 ml volumetric flask up to the scale
line. Mix the solution until the reagent is dissolve.
2. Do the same for the pH 4.01 and pH 10.01 packets.
3. You will need only a small amount of calibration solution during calibration. Get the require solution from the flask
and store the rest.
4. Do not reuse used solutions.
SETUP MENU
The KS-PH-221-DELUXE has integrated set-up menus for each of the parameters it can measure. Below is a guide
showing the available options for each parameter, the available options and their description.
pH MODE
PARAMETER
DESCRIPTION
OPTIONS
DESCRIPTION
BUF
pH buffers:
Select your desired pH buffer for
calibration and auto-recognition
USA (Default)
USA (1.68, 4.01, 7.00,
10.01, 12.45)
NIST
NIST (1.68, 4.01, 6.96,
9.18, 12.45)
DIN
DIN (1.09, 4.65, 6.79, 9.23,
12.75)
USER
User-Defined Buffer
CAL
Calibration Points:
Select the desired number of
calibration points
1
1 point
2
2 points
3 (Default)
3 points
4
4 points
5
5 points
RESO
Resolution:
Sets the resolution for pH
measurement
0.001 (Default)
0.001 pH
0.01
0.01 pH
0.1
0.1 pH
UNIT
Measurement unit:
Sets the default temperature unit
°C (Default)
Degrees Celsius
°F
Degrees Fahrenheit
ORP MODE
PARAMETER
DESCRIPTION
OPTIONS
DESCRIPTION
RESO
Resolution:
Sets the resolution for pH
measurement
0.1
0.1 mV
1
1 mV