background image

5

through the condenser for a standard microscope slide.  Depressing the
left side switch turns on the supra-stage light for macro and opaque
objects.  

Built-in Illuminator(s)

T-1924C - There are two illuminators built into to each of the microscope

bases, one is sub-stage and one is supra-stage.  The supra-stage
illuminator is necessary for macro or opaque specimens where light
from the sub-stage illuminator can not come through.  Only one of
these illuminators can be on at a time (not both), but both will change
their intensity by using the rheostat.

Focus

Be sure the X-axis (in the middle of the Bridge Assembly) control

knob lever is in the "unlock" position, and while looking through the
eyepieces make sure the X-axis control knob is in the middle.  Each
microscope works independently and must be focused independently.
Note - In the T-1924C, be sure that the floating stage is centered directly
over the sub-stage illuminator, so light comes through. - The floating
stage can actually cover up the sub-stage illuminator so no light comes
through.

Place a specimen slide on the stage of the microscope, and move the

specimen into place for viewing.  First, put the 4x objective in position of
use.  Using the coarse adjust knob (larger of two) raise the stage until it
will go no higher.  Now looking through the eyepieces, lower the stage
(again using the coarse adjust), bringing the specimen into focus.  Use
the fine adjust knob (smaller of two) to now sharpen the focus and attain
maximum resolution.  Now rotate the 10x objective into place, and you
should only have to use the fine adjust to achieve a crisp focus (The
microscope is Parfocal).  Now the 40x objective can be similarly utilized if
needed.  

Note - on the T-1924C you may start with the 2x objective if you choose
to.

With coaxial focusing, both the coarse focus (large knob) and fine

focus (small knob) are on the same axis.  Having both sets of knobs on
the same axis makes it easier to switch from one focus knob to the other
without moving eyes away from the eyepieces.  Repeat this procedure
with each of the two microscope bases.

Summary of Contents for Comparison Scope T-1924C

Page 1: ...Cordless Comparison Scope Instruction Manual T 1924C...

Page 2: ...or opaque objects in a manner similar to a stereo or dissecting microscope Both are shipped as cordless units with an enclosed charger power cord adapter The T 1924C has 2x 4x 10x and 40xR DIN objecti...

Page 3: ...ft each microscope base out of shipping container Note that the serial numbers on the bottom of the units end in either an L left base or an R right base Remove all packing materials on stage and illu...

Page 4: ...he socket sliding in from the right until it is flush in socket opening Tighten the set screw Step 6 Insert eyepieces into the eyetubes Step 7 Plug both microscopes into wall socket and allow them to...

Page 5: ...comes through The floating stage can actually cover up the sub stage illuminator so no light comes through Place a specimen slide on the stage of the microscope and move the specimen into place for vi...

Page 6: ...ual s eyes The Seidentopf binocular head allows adjustment for interpupillary interocular adjustment The eyepieces will move to exactly match the distance between the user s eyes Care and Maintenance...

Page 7: ...el style center pin with a 5 5 volt 300mA output supply which may be used with 110v 220v 230v or 240v input systems The power supply charger is made exclusively for the Ken A Vision Cordless Microscop...

Page 8: ...Y WARRANTY To replace the batteries loosen the silver thumbscrew in bottom of the base Drop the door down remove the battery cover and replace the batteries Accessories If you are interested in purcha...

Reviews: