The IODDs required for configuring the IO-Link device as well as detailed information on process data
setup, diagnostic functions and parameter addresses are available in the download area at
www.keller.de/its.
ATTENTION
For IO-Link operation, a 3-wire shielded cable port Class A (Type A) must
be used.
12 Trouble Shooting
No measured values are determined.
The measuring system CellaCast checks the validity of the measured values at each measure-
ment and calculates the signal strength. The value for the signal strength must be > 5 %. The
signal strength is only indicated during the casting process.
Parameter
Calculated signal strength
Q.Si6
In [%]
Checking the signal intensity
► Press [Mode]
> The display shows [
Ao
]
► Press the key [▼] until the parameter [
Qu
] is shown in the display
► Press [Mode]
> The display shows [
Q.EPS
]
► Press the key [▼] until the parameter [
Q.Si6
] is shown
► Press [Mode]
> The signal strength is indicated in the display.
If the displayed signal strength is < 5 %, perform the following checks:
Check the lens or protective screen for contamination.
Check the adjustment and focusing.
• The pyrometer must be focused on the measuring distance.
• The measuring field must be filled to 30 %.
• The pyrometer must have free view to the pouring stream.
13 Shipping, packing and disposal
Inspection after shipping
Unpack and inspect the entire shipment immediately upon receipt to make sure it is complete
and undamaged. If the container/package shows visible signs of damage, please refuse the
shipment. If this is not possible, accept the shipment on the condition that the freight carrier’s de-
livery record is noted with the extent of the damage in order to file a claim. Should you discover
a concealed loss or damage, report it to the shipper or freight carrier immediately. If the period
for filing claims has expired, you will no longer be able to make any claims for compensation
of damage or loss.
Quick Start Guide CellaCast PX 8x
Summary of Contents for ITS CellaCast PX 8 Series
Page 1: ...CellaCast PX 8x 1109371 12 2020...
Page 2: ......
Page 53: ...RU CellaCast PX 8x 1 X 8x CellaCast PX www keller de its 2 3 2 1 0 630 680 4 CellaCast PX 8x...
Page 54: ...RU 1 5 30 30 50 6 65 C 1 3 2 4 CellaCast PX 8x...
Page 56: ...RU 8 PX xx AF xx C PX xx AF xx L 2 IO Link 2 IO Link DIN 916 DIN 911 1 CellaCast PX 8x...
Page 60: ...RU 1 2 C 101 104 2 103 0 IO Link CellaCast PX 8x...
Page 62: ...RU 13 KELLER HCW IO Link CellaCast PX 8x...
Page 63: ...CN CellaCast PX 8x 1 PX 8x www keller de its 2 3 1 0mW 630 670nm 4 CellaCast PX 8x...
Page 64: ...CN 1 5 30 30 50 6 65 C 1 3 2 4 CellaCast PX 8x...
Page 66: ...CN PX xx AF xx C PX xx AF xx L 2 IO Link 2 IO Link DIN 911 DIN 916 1 60 C LED 2 15...
Page 69: ...CN Mode Ao Qu Mode Q EPS Q ATD Mode t dEl t Act Mode 2 3 Mode 30 End End IO Link 10...
Page 70: ...CN 1 2 C 101 104 103 0 IO Link 11 IO IO Link IO Link IO Link IO Link...
Page 72: ...CN KELLER HCW IO Link CellaCast PX 8x...
Page 74: ...KR 1 5 30 30 6 bar 50 l min 65 C 1 Target size 3 field of view 2 4 CellaCast PX 8x...
Page 80: ...KR CellaCast PX 8x 1 Ratio correction 2 C 101 104 2 103 0 IO Link 11 IO IO IO IO IO...
Page 82: ...KR KELLER HCW IO Link CellaCast PX 8x...
Page 83: ......