EN
1: Ideal measuring point
5 Mechanical installation
The distance should be adjusted in that way, that the pouring stream is wide enough to fill up
at least 30 % of the measurement area. Reduce the distance, if the pouring stream covers less
than 30 % of the measurement area. When aiming the pyrometer, make sure that the pouring
stream is in the centre of the measurement area. During pouring it must be ensured that the
pouring stream is always within the measuring field, even if the pouring stream is fluctuating.
In order to prevent the pyrometer optics from contamination, connect the scavenging air. Ap-
prox. 50 l/min with a maximum pressure of 6 bars are required for that. Use oil-free and dirt-free
compressed air. If the expected ambient temperatures are > 65 °C, a cooling system is required.
1: Target size
3: Field of view
2: Pouring stream
4: Distance to target
Quick Start Guide CellaCast PX 8x
Summary of Contents for ITS CellaCast PX 8 Series
Page 1: ...CellaCast PX 8x 1109371 12 2020...
Page 2: ......
Page 53: ...RU CellaCast PX 8x 1 X 8x CellaCast PX www keller de its 2 3 2 1 0 630 680 4 CellaCast PX 8x...
Page 54: ...RU 1 5 30 30 50 6 65 C 1 3 2 4 CellaCast PX 8x...
Page 56: ...RU 8 PX xx AF xx C PX xx AF xx L 2 IO Link 2 IO Link DIN 916 DIN 911 1 CellaCast PX 8x...
Page 60: ...RU 1 2 C 101 104 2 103 0 IO Link CellaCast PX 8x...
Page 62: ...RU 13 KELLER HCW IO Link CellaCast PX 8x...
Page 63: ...CN CellaCast PX 8x 1 PX 8x www keller de its 2 3 1 0mW 630 670nm 4 CellaCast PX 8x...
Page 64: ...CN 1 5 30 30 50 6 65 C 1 3 2 4 CellaCast PX 8x...
Page 66: ...CN PX xx AF xx C PX xx AF xx L 2 IO Link 2 IO Link DIN 911 DIN 916 1 60 C LED 2 15...
Page 69: ...CN Mode Ao Qu Mode Q EPS Q ATD Mode t dEl t Act Mode 2 3 Mode 30 End End IO Link 10...
Page 70: ...CN 1 2 C 101 104 103 0 IO Link 11 IO IO Link IO Link IO Link IO Link...
Page 72: ...CN KELLER HCW IO Link CellaCast PX 8x...
Page 74: ...KR 1 5 30 30 6 bar 50 l min 65 C 1 Target size 3 field of view 2 4 CellaCast PX 8x...
Page 80: ...KR CellaCast PX 8x 1 Ratio correction 2 C 101 104 2 103 0 IO Link 11 IO IO IO IO IO...
Page 82: ...KR KELLER HCW IO Link CellaCast PX 8x...
Page 83: ......