EN
Quick Start Guide CellaCast PX 8x
6 Assembly fibre optic cable
One end of the optical fibre has a nameplate showing the serial number of the corresponding
basic pyrometer. This is the end which must be screwed onto the pyrometer. For proper connec-
tion, the arrow on the nameplate of the fibre optic cable and the arrow on the pyrometer should
point toward each other. The serial number of the measuring head should also correspond to
the pyrometer.
ATTENTION
The fibre optic cable must not be exposed to tensile load and must not be
twisted. The minimum bending radius is 125 mm.
7 Electrical connection
The pyrometer is supplied with low voltage 24 DC (18 ... 32 V DC).
ATTENTION
The device may only be installed by a skilled, qualified electrician. Do not
connect it while the voltage supply source is turned on. Please observe
international safety regulations at all times.
►
Disconnect the pyrometer from any voltage source.
►
Connect the pyrometer according to the following schematic.
Pin 1 BN (brown)
L+ (Power supply24V DC)
Pin 4 BK (black)
Open Collector switching output; I
max
= 150 mA or IO-Link OUT 1
Pin 5 GY (grey)
Open Collector switching output; I
max
= 150 m OUT2
Pin 2 WH (white)
Analogue output; 0/4 ... 20mA
Pin 3 BU (blue)
L- (GND)
The pyrometer must be protected against electromagnetic fields. Therefore, use a shiel-
ded cable, connecting it via connector casing of the pyrometer to the housing.
Use a flyback diode when switching inductive loads.
Summary of Contents for ITS CellaCast PX 8 Series
Page 1: ...CellaCast PX 8x 1109371 12 2020...
Page 2: ......
Page 53: ...RU CellaCast PX 8x 1 X 8x CellaCast PX www keller de its 2 3 2 1 0 630 680 4 CellaCast PX 8x...
Page 54: ...RU 1 5 30 30 50 6 65 C 1 3 2 4 CellaCast PX 8x...
Page 56: ...RU 8 PX xx AF xx C PX xx AF xx L 2 IO Link 2 IO Link DIN 916 DIN 911 1 CellaCast PX 8x...
Page 60: ...RU 1 2 C 101 104 2 103 0 IO Link CellaCast PX 8x...
Page 62: ...RU 13 KELLER HCW IO Link CellaCast PX 8x...
Page 63: ...CN CellaCast PX 8x 1 PX 8x www keller de its 2 3 1 0mW 630 670nm 4 CellaCast PX 8x...
Page 64: ...CN 1 5 30 30 50 6 65 C 1 3 2 4 CellaCast PX 8x...
Page 66: ...CN PX xx AF xx C PX xx AF xx L 2 IO Link 2 IO Link DIN 911 DIN 916 1 60 C LED 2 15...
Page 69: ...CN Mode Ao Qu Mode Q EPS Q ATD Mode t dEl t Act Mode 2 3 Mode 30 End End IO Link 10...
Page 70: ...CN 1 2 C 101 104 103 0 IO Link 11 IO IO Link IO Link IO Link IO Link...
Page 72: ...CN KELLER HCW IO Link CellaCast PX 8x...
Page 74: ...KR 1 5 30 30 6 bar 50 l min 65 C 1 Target size 3 field of view 2 4 CellaCast PX 8x...
Page 80: ...KR CellaCast PX 8x 1 Ratio correction 2 C 101 104 2 103 0 IO Link 11 IO IO IO IO IO...
Page 82: ...KR KELLER HCW IO Link CellaCast PX 8x...
Page 83: ......