Operating instructions CellaTemp PX 4x 5x 6x
The pyrometer´s field of view must remain unobstructed� Any interference from
objects must be avoided�
ATTENTION
At an ambient temperature > 65 °C, the pyrometer must be cooled or protected
against radiation by a shielding plate�
7.2 Alignment
Align the pyrometer with the object to be measured� The pyrometer should, if pos-
sible, be installed at an angle of 90 ° to the measuring object� The angle should not
be less than 45 ° from the vertical� When measuring with a two-colour pyrometer,
the field of view does not have to be filled entirely by the target object�
1 Radiated energy
2 Visibly smaller than measuring spot
3 Lens or protective screen contaminated
4 Steam, dirt or gas in the atmosphere
5 Object is smaller than the measuring field or object is moving
If the pyrometer is operated in spectral measurement mode, the measure-
ment object must completely fill the measuring field�
1
1
1
1
2
3
4
5