30
11.2 One-colour mode
►
Set the pyrometer to one-colour mode:
One-colour [
Ao
] → [
Ao.S
] =
SpEc.
►
Align the pyrometer to the object to be measured
The pyrometer uses the intensity of infrared radiation for non-contact temperature
measurements� It is necessary to configure the pyrometer to the respective
emissivity coefficient of the measuring object to obtain exact measuring results (→
15 Emissivity coefficient tables)� An incorrectly set emissivity coefficient leads to
wrong temperature readings�
Set the emissivity coefficient after connecting the supply voltage or resetting the
parameters to factory settings�
►
Press [
▲or ▼]
> the value of the selected emissivity is displayed, for example [
100.0
]
►
Press [
▲or ▼] until the desired emissivity is shown
►
Press [Enter] or wait for 3 seconds
The current temperature value is displayed� The pyrometer now works with this
configured emissivity until it is changed again�
Target larger
than spot size
Best
Good
Incorrect
Target equal
to spot size
Target smaller
than spot size