background image

Models 2601B, 2602B and 2604B System SourceMeter

®

 Instrument Specifications 

 

Specifications are subject to change without notice 

12 of 13 

SPEC-2602B Rev. B / February 2014 

GENERAL SPECIFICATIONS 

IEEE-488 

IEEE Std 488.1 compliant. Supports IEEE Std 488.2 common commands and status 
model topology 

RS-232 

 

Baud rates from 300 bps to 115,200 bps 

 

Programmable number of data bits, parity type, and flow control (RTS/CTS hardware 
or none) 

 

When not programmed as the active host interface, the Models 2601B, 2602B and 
2604B can use the RS-232 interface to control other instruments 

Ethernet 

RJ-45 connector, LXI version 1.4 Core 2011, 10/100BaseT, Auto-MDIX 

LXI compliance 

LXI version 1.4 Core 2011

 

Expansion interface

24

 

 

The TSP-Link

®

 expansion interface allows TSP-enabled instruments to trigger and 

communicate with each other 

 

Cable type: Category 5e or higher LAN crossover cable 

 

9.84 ft (3 m) maximum between each TSP-enabled instrument

 

USB Control (Rear) 

USB 2.0 Device:  USB-TMC488 protocol 

USB File System (Front) 

USB 2.0 Host:  Mass storage class device 

Power supply 

100 V to 240 V AC, 50 Hz or 60 Hz (auto sensing), 240 VA maximum 

Cooling 

Forced air; side intake and rear exhaust.  One side must be unobstructed when rack 
mounted. 

Warranty 

1 year 

EMC 

Conforms to European Union EMC Directive 

Safety 

NRTL listed to UL61010-1:2008 and CSA C22.2 No. 61010-1 

Conforms to European Union Low Voltage Directive 

Environment 

For indoor use only 

Altitude:

 Maximum 6562 ft (2000 m) above sea level 

Operating:

 0 °C to 50 °C, 70 % relative humidity up to 35 °C. Derate 3 % relative 

humidity/°C, 35 °C to 50 °C 

Storage:

 -25 °C to 65 °C 

Dimensions 

Rack mount: 3.5 in. high × 8.4 in. wide × 17.5 in. deep (89 mm × 213 mm × 460 mm)  

Bench configuration (with handle and feet): 4.1 in. high × 9.4 in. wide × 17.5 in. deep 
(104 mm × 238 mm × 460 mm) 

Weight 

2601B:

 10.4 lb  (4.75 kg) 

2602B and 2604B:

 12.0 lb (5.50 kg) 

                                              

24

 TSP-Link is not available with the Model 2604B. 

Summary of Contents for System SourceMeter 2601B

Page 1: ...and measurement accuracies are specified at the Models 2601B 2602B and 2604B terminals under these conditions 1 23 C 5 C 70 percent relative humidity 2 After a two hour warm up period 3 Speed normal...

Page 2: ...pA 0 03 5 nA 60 pA 10 pA 0 025 1 5 nA 100 A 2 nA 0 03 60 nA 3 nA 100 pA 0 02 25 nA 1 mA 20 nA 0 03 300 nA 6 nA 1 nA 0 02 200 nA 10 mA 200 nA 0 03 6 A 200 nA 10 nA 0 02 2 5 A 100 mA 2 A 0 03 30 A 600...

Page 3: ...ibe the typical performance of the Models 2601B 2602B and 2604B PULSE CHARACTERISTICS Pulse region specifications Pulse region specifications Region quadrant diagram Region maximums Maximum pulse widt...

Page 4: ...HARACTERISTICS Noise 10 Hz to 20 MHz 20 mV peak peak 3 mV RMS 6 V range Transient response time 70 s for the output to recover to within 0 1 for a 10 to 90 step change in load Overshoot Voltage 0 1 10...

Page 5: ...nal value after source level command is processed on a fixed range Voltage range Settling time 100 mV 50 s 1 V 50 s 10 V 110 s 40 V 12 150 s Current source output settling time Time required to reach...

Page 6: ...quired to reach within 0 1 of final value after source level command is processed on a fixed range Values below for Vout 1 V Current range Settling time 1 mA 100 s Input impedance 10 G ADDITIONAL CHAR...

Page 7: ...ge Settling time 3 A and 1A 120 s Rload 2 100 mA and 10 mA 100 s 1 mA 3 ms 100 A 3 ms 10 A 230 ms 1 A 230 ms Capacitor leakage performance Using HIGH C scripts 19 200 ms 50 nA Load 5 F in parallel wit...

Page 8: ...01 NPLC Digital I O 3500 3100 3400 3000 3000 2700 2900 2600 4150 3650 3800 3400 0 1 NPLC Internal 580 480 560 470 550 465 550 460 560 470 545 460 0 1 NPLC Digital I O 550 460 550 460 540 450 540 450 5...

Page 9: ...maximum rate is 190 second Maximum source function change rate 1000 per second Command processing time 1 ms Maximum time required for the output to begin to change after receiving the smua source leve...

Page 10: ...ts for example branching looping and math Able to execute high speed test scripts stored in memory without host intervention Minimum user memory available 16 MB approximately 250 000 lines of TSP code...

Page 11: ...ink expansion interface allows TSP enabled instruments to trigger and communicate with each other See the figure below Each Model 2601B and 2602B has two TSP Link connectors to make it easier to conne...

Page 12: ...or higher LAN crossover cable 9 84 ft 3 m maximum between each TSP enabled instrument USB Control Rear USB 2 0 Device USB TMC488 protocol USB File System Front USB 2 0 Host Mass storage class device...

Page 13: ...current flowing out of digital I O bit 960 A Maximum sink current at maximum logic low voltage 0 7 5 0 mA Absolute maximum sink current flowing into digital I O pin 11 mA 5 V power supply pins Limited...

Reviews: