2600S-901-01 Rev. C / January 2008
Return to
14-3
Series 2600 System SourceMeter® Instruments Reference Manual
Section 14: Display Operations
2
or
display.SMUA_SMUB
3
or
display.USER
Display screen example:
The following command displays source-measure and compliance for SMU A:
display.screen = display.SMUA
Measurement functions
With a source-measure screen selected, the measured reading can be displayed as volts, amps,
ohms or watts.
The
display.smuX.measure.func
attribute is used to select the displayed measure function:
display.smuX.measure.func = function
where:
smuX
=
smua
or
smub
function
is set to one of the following values:
0
or
display.MEASURE_DCAMPS
1
or
display.MEASURE_DCVOLTS
2
or
display.MEASURE_OHMS
3
or
display.MEASURE_WATTS
Measurement function example:
The following command sets SMU A to display ohms measurements:
display.smua.measure.func = display.MEASURE_OHMS
Display resolution
Display resolution for measured readings can be set to 4-1/2, 5-1/2 or 6-1/2 digit resolution.
The
display.smuX.digits
attribute is used to set display resolution for measured readings:
display.smuX.digits = digits
Where:
smuX
=
smua
or
smub
digits
is set to one of the following values:
4
or
display.DIGITS_4_5
5
or
display.DIGITS_5_5
6
or
display.DIGITS_6_5
Display resolution example:
The following command sets SMU A for 5-1/2 digit resolution for measured readings:
display.smua.digits = display.DIGITS_5_5