Table 2-4
Front panel current calibration
Source range
1
Source current
Multimeter
current reading
2
10
μ
A
+10.000
μ
A
+00.000
μ
A
-10.000
μ
A
+00.000
μ
A
___________
μ
A
___________
μ
A
___________
μ
A
___________
μ
A
100
μ
A
+100.00
μ
A
+000.00
μ
A
-100.00
μ
A
+000.00
μ
A
___________
μ
A
___________
μ
A
___________
μ
A
___________
μ
A
1mA
+1.0000mA
+0.0000mA
-1.0000mA
+0.0000mA
___________ mA
___________ mA
___________ mA
___________ mA
10mA
+10.000mA
+00.000mA
-10.000mA
+00.000mA
___________ mA
___________ mA
___________ mA
___________ mA
100mA
+100.00mA
+000.00mA
-100.00mA
+000.00mA
___________ mA
___________ mA
___________ mA
___________ mA
1A
+1.0000A
+0.0000A
-1.0000A
+0.0000A
___________ A
___________ A
___________ A
___________ A
3A
3
+3.0000A
+0.0000A
-3.0000A
+0.0000A
___________ A
___________ A
___________ A
___________ A
1
Use EDIT and RANGE keys to select source range.
2
Multimeter reading used in corresponding calibration step. See procedure.
3
Current calculated as follows: I=V/R, where V is the DMM reading, and R is the
characterized value of the 1
Ω
resistor.
2-16
Calibration
Summary of Contents for SourceMeter 2425
Page 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031...
Page 15: ......
Page 16: ...1 PerformanceVerification...
Page 35: ...1 20 Performance Verification Model 2425 Service Manual...
Page 36: ...2 Calibration...
Page 62: ...3 RoutineMaintenance...
Page 65: ...3 4 Routine Maintenance Model 2425 Service Manual...
Page 66: ...4 Troubleshooting...
Page 84: ...5 Disassembly...
Page 92: ...6 ReplacementParts...
Page 113: ...6 22 Replacement Parts Model 2425 Service Manual...
Page 114: ...A Specifications...
Page 116: ...B CommandReference...
Page 128: ...C CalibrationPrograms...
Page 139: ......
Page 141: ......