Calibration
2-7
Changing the password
The default password may be changed from the front panel or via remote as discussed
next.
Changing the password from the front panel
Follow the steps below to change the password from the front panel:
1. Press the MENU key, then choose CAL and press ENTER. The instrument will
display the following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES
©
3
SAVE LOCK CHANGE-PASSWORD
2. Select UNLOCK, then enter the password. (Default: 002425.)
3. Select CHANGE-PASSWORD, then press ENTER. The instrument will display
the following:
New Pwd: 002425
Use
ß
,
©
,
¹
,
ƒ
,
ENTER or EXIT.
4. Using the range keys and the left and right arrow keys, enter the new password on
the display.
5. Once the desired password is displayed, press the ENTER key to store the new
password.
Changing the password by remote
To change the calibration password by remote, first send the present password, and then
send the new password. For example, the following command sequence changes the pass-
word from the 'KI002425' remote default to 'KI_CAL':
:CAL:PROT:CODE 'KI002425'
:CAL:PROT:CODE 'KI_CAL'
You can use any combination of letters and numbers up to a maximum of eight characters.
NOTE
If you change the first two characters of the password to something other than
“KI”, you will not be able to unlock calibration from the front panel.
Summary of Contents for SourceMeter 2425
Page 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031...
Page 15: ......
Page 16: ...1 PerformanceVerification...
Page 35: ...1 20 Performance Verification Model 2425 Service Manual...
Page 36: ...2 Calibration...
Page 62: ...3 RoutineMaintenance...
Page 65: ...3 4 Routine Maintenance Model 2425 Service Manual...
Page 66: ...4 Troubleshooting...
Page 84: ...5 Disassembly...
Page 92: ...6 ReplacementParts...
Page 113: ...6 22 Replacement Parts Model 2425 Service Manual...
Page 114: ...A Specifications...
Page 116: ...B CommandReference...
Page 128: ...C CalibrationPrograms...
Page 139: ......
Page 141: ......