Command Reference
B-5
:DATE
(:CALibration:PROTected:DATE)
Purpose
To program the calibration date.
Format
:cal:prot:date <year>, <month>, <day>
Parameters
<year> = 1994 to 9999
<month> = 1 to 12
<day> = 1 to 31
Query
:cal:prot:date?
Response
<year> , <month> , <day>
Description
The :DATE command allows you to store the calibration date in instrument
EEROM for future reference. You can read back the date from the instru-
ment by using the :DATE? query, or by using the front panel CAL menu.
Note
The year, month, and day parameters must be delimited by commas.
Example
:CAL:PROT:DATE 2000,7,20
Send cal date (7/20/00).
:NDUE
(:CALibration:PROTected:NDUE)
Purpose
To send the next calibration due date to the instrument.
Format
:cal:prot:ndue <year>, <month>, <day>
Parameters
<year> = 1994 to 9999
<month> = 1 to 12
<day> = 1 to 31
Query
:cal:prot:ndue?
Response
<year>, <month>, <day>
Description
The :NDUE command allows you to store the date when calibration is next
due in instrument memory. You can read back the next due date by using the
:NDUE? query, or by using the front panel CAL menu.
Note
The next due date parameters must be delimited by commas.
Example
:CAL:PROT:NDUE 2001,11,20
Send due date (11/20/01).
Summary of Contents for SourceMeter 2425
Page 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031...
Page 15: ......
Page 16: ...1 PerformanceVerification...
Page 35: ...1 20 Performance Verification Model 2425 Service Manual...
Page 36: ...2 Calibration...
Page 62: ...3 RoutineMaintenance...
Page 65: ...3 4 Routine Maintenance Model 2425 Service Manual...
Page 66: ...4 Troubleshooting...
Page 84: ...5 Disassembly...
Page 92: ...6 ReplacementParts...
Page 113: ...6 22 Replacement Parts Model 2425 Service Manual...
Page 114: ...A Specifications...
Page 116: ...B CommandReference...
Page 128: ...C CalibrationPrograms...
Page 139: ......
Page 141: ......