![background image](http://html1.mh-extra.com/html/keithley/sourcemeter-2400-series/sourcemeter-2400-series_quick-results-manual_4084871024.webp)
16
Quick Results Guide
SCPI commands
SCPI commands to source-measure and measure only are provided in Table 6, and the com-
mands to measure ohms are provided in Table 7.
Table 6
SCPI commands; source-measure and measure only
1
Command
2
Description
:SOURce:FUNCtion <name>
Select source function; <name> = VOLTage or CURRent.
:SOURce:xxx:MODE FIXed
Select fixed V or I sourcing mode.
:SOURce:xxx:RANGe <n>
Select measurement range for V or I source; <n> = range.
:SOURce:xxx:LEVel <n>
Set source amplitude; <n> = amplitude in volts or amps.
[:SENSe]:FUNCtion <name>
Select measure function; <name> = “VOLTage” or “CURRent”.
[:SENSe]:xxx:PROTection <n>
Set V or I compliance; <n> = compliance.
[:SENSe]:xxx:RANGe <n>
Select V or I measure range; <n> = range.
[:SENSe]:xxx:RANGe:AUTO <b> Enable or disable auto range; <b> = ON or OFF.
:OUTPut <b>
Turn output on or off; <b> = ON or OFF.
:READ?
Trigger and acquire one data string.
1. For measure only, the following rules apply (Table 9 provides an example to measure I only):
a.
To measure V only, you must source I. To measure I only, you must source V.
b.
The source must be set to 0V or 0A.
c.
When measuring V only, DO NOT use auto range.
d.
When measuring V only, voltage compliance MUST be set higher than the voltage being measured.
2. :xxx = :VOLTage or :CURRent.
Table 7
SCPI commands; measure ohms
Command Description
[:SENSe]:FUNCtion “RESistance”
Select ohms function.
[:SENSe]:RESistance:MODE <name>
Select ohms mode; <name> = AUTO or MANual*.
[:SENSe]:RESistance:OCOMpensated <b> Enable/disable offset compensation.
[:SENSe]:RESistance:RANGe <n>
Select range for auto ohms; <n> = range.
[:SENSe]:RESistance:RANGe:AUTO <b>
Enable or disable auto range for auto ohms; <b> = ON or
OFF.
:OUTPut <b>
Turn output on or off; <b> = ON or OFF.
:READ?
Trigger and acquire one data string.
*For manual ohms, you must configure the SourceMeter to source-measure. Table 11 provides a programming example to
measure ohms using the manual ohms method.