*SAV
This command saves the present instrument settings as a user-saved setup.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Nonvolatile memory
Not applicable
Usage
*SAV <n>
<n>
An integer from 0 to 4
Details
Save the present instrument settings as a user-saved setup. You can restore the settings with the
*RCL
command.
Any command that is affected by
*RST
can be saved with the
*SAV
command.
Any settings that had been stored previously as
<n>
are overwritten.
If you view the user-saved setups from the front panel of the instrument, they are stored as scripts named
Setup0<n>
.
Example
*SAV 2
Saves the instrument settings in memory location
2.
Also see
(on page 6-1)
:ABORt
This command stops all trigger model commands on the instrument.
Type
Affected by
Where saved
Default value
Command
Not applicable
Not applicable
Not applicable
Usage
:ABORt
Details
When this command is received, the Model 2450 stops the trigger model.
Also see
(on page 3-108)
(on page 3-95)