background image

The graph in the 

Figure 6

 shows the results of this 

application. Notice the exponential current response of the 
capacitor as it charges up to 20 V over time.

Figure 6. Leakage current test results.

Measuring insulation resistance

The following applications demonstrate how to use the Model 
2450 to measure insulator resistance. The applications shows 
how to use the front-panel interface, the remote interface using 
the SCPI command set, and the remote interface using the Test 
Script Processor (TSP

®

) command set.

You make insulation resistance measurements between traces 

on a printed circuit board and between conductors in cables and 
connectors.

This application sets the Model 2450 to source 20 V and 

makes 10 resistance readings with a 100 ms interval time. Once 
the measurements are made, the output is turned off.

For this test, you will:

•  Reset the instrument
•  Set the instrument to read the rear terminals
•  Select the source voltage function and measure 

resistance function

•  Set the magnitude of the voltage source output
•  Turn on autoranging
•  Use the Simple Loop trigger model template to set the 

numbers of readings and interval time

•  Turn on the source output
•  Make readings
•  Turn off the source output

Set up the insulation resistance 

application using the front panel

To set up the application from the front panel:

1.  Connect the device under test (DUT) to the rear panel of the 

Model 2450, as described in 

Device connections

.

2.  Reset the Model 2450.
3.  Press the 

MENU

 key.

4.  Under System, select 

Manage

.

5.  Select 

System Reset

, and then select 

OK

.

6.  Press the 

TERMINAL FRONT/REAR

 switch to set the 

instrument to use the rear-panel terminals (R is displayed to 
the left of the switch).

7.  Press the 

HOME

 key.

8.  Press the 

FUNCTION

 key.

9.  Under Source Voltage and Measure, select 

Resistance

.

10. Select the button next to Source (at the bottom of the screen).
11. Enter 

20 V

 (or other applicable test voltage) and select 

OK

.

12. Press the 

MENU

 key.

13. Under Trigger, select 

Templates

.

14. Next to Templates, select 

SimpleLoop

.

15. Set Count to 

10

 and select 

OK

.

16. Set Delay to 

0.1

 seconds and select 

OK

.

17. Press the 

HOME

 key.

18. Press the 

OUTPUT ON/OFF

 switch to enable the output.

19. Press the 

TRIGGER

 key to begin taking readings. The 

resistance measurements appear in the measure area (top 
half) of the Home screen.

20. Press the 

OUTPUT ON/OFF

 switch to turn off the output 

when you are done making measurements.

-- Reset the instrument, which also clears the buffer. 
reset()
-- Set up the source function. smu.source.func = smu.
FUNC_DC_VOLTAGE smu.source.ilimit.level = 10e-3 smu.
source.level = 20
-- Set up measure function. smu.measure.func = smu.
FUNC_DC_CURRENT smu.measure.terminals = smu.TERMINALS_
REAR smu.measure.autorange = smu.ON smu.measure.nplc = 
1
-- Turn on the output and initiate readings. 
smu.source.output = smu.ON trigger.model.
load(“DurationLoop”, 60, 0.2) trigger.model.initiate()
-- Wait for the trigger model to complete. 
waitcomplete()
-- Parse index and data into three columns. print(“Rdg 
#”, “Time (s)”, “Current (A)”) for i = 1, defbuffer1.n 
do
print(i, defbuffer1.relativetimestamps[i], 
defbuffer1[i])
end
-- Discharge the capacitor to 0 V and turn off the 
output. smu.source.level = 0
delay(2)
smu.source.output = smu.OFF