Figure 3. Rear-panel connections for the leakage current test.

Figure 4. Rear-panel connections insulation resistance test.

Measuring leakage current

The following application demonstrates how to use the Model 
2450 to measure the leakage current of a 1nF capacitor by 
sourcing a voltage and measuring the resulting current using the 
front panel or over the remote interface. The remote interface 
examples show SCPI commands and TSP commands.

This application sets the Model 2450 to source 20V and 

measure the resulting leakage current as a function of time. The 
instrument takes current measurements for a specific period.

For this test, you will:

•  Reset the instrument.
•  Set the instrument to read the rear terminals.
•  Select the source voltage function and measure 

current function.

•  Set the magnitude of the voltage source.
•  Turn on autoranging.
•  Set the measure delay.
•  Use the Duration Loop trigger model template to make 

readings for a specified time period.

•  Turn on the source output.
•  Take readings for a specified period of time.
•  Turn off the source output.

NOTE:

 When you use the Model 2450 to measure small current 

values, ensure that the device under test is electrostatically 
shielded. If the capacitor rating is greater than 20nF, 
enable the high capacitance mode for best results.

 

For more information about making optimized 
capacitor leakage measurements and minimizing noisy 
measurements, see the Keithley Instruments 

Low Level 

Measurements Handbook

, available on the Keithley 

Instruments website (

http://www.keithley.com

).

Set up the leakage current application 
using the front panel

To set up the application from the front panel:

1.  Use the test leads to make the connection from the capacitor 

to the rear panel of the Model 2450, as described in 

Device connections

.

2.  Reset the Model 2450.

a. Press the 

MENU

 key.

b. Under System, select 

Manage

.

c. Select 

System Reset

, and then select 

OK

.

3.  Press the 

TERMINAL FRONT/REAR

 switch to set the 

instrument to use the rear-panel terminals (R is displayed to 
the left of the switch).

4.  Press the 

HOME

 key.

5.  Press the 

FUNCTION

 key.

6.  Under Source Voltage and Measure, select 

Current

.

7.  Select the button next to Source (at the bottom of the screen).
8.  Enter 

20 V

 and select 

OK

.

9.  Press the 

MENU

 key.

10. Under Trigger, select 

Templates

.

11. Next to Templates, select 

Duration Loop

.

12. Next to Duration, enter a soak time of 

30s

 and select 

OK

.

13. Next to Delay, enter 

0.2s

, and select 

OK

.

14. Press the 

HOME

 key to return to the Home screen.

15. Press the 

TRIGGER

 key to turn on the output and 

make measurements. The output is turned off when the 
measurements are complete.

Figure 5. Viewing the leakage current measurement on the front-panel.