Connections and Installation
2-3
2.3.4 Typical connecting scheme
Figure 2-3 shows a typical connecting scheme for the Mod-
el 7017. In this arrangement, sources are connected to the
inputs, and the measuring instrument is connected to the
outputs.
CAUTION
Maximum voltage from any terminal
(center conductor or shield) to any other
terminal or chassis is 42V peak. Exceed-
ing this value may result in card damage.
Figure 2-3
Typical connection scheme
IN 1
IN 2
OUT A
IN 3
IN 4
IN 5
IN 6
OUT B
IN 7
IN 8
Note: All cables are 50
Ω
Measuring Instrument
AC
Signals
2.3.3 Input/output connecting cables
All connections to the multiplexer card input and output
jacks should be made using 50
Ω
coaxial cable equipped with
SMA connectors. RG223/U is the recommended cable type.
WARNING
Make sure power is off and external en-
ergy sources are discharged before con-
necting or disconnecting cables.
NOTE
50
Ω
cables must be used to ensure good
high-frequency performance. RG223/U ca-
bles should be used for best performance.
See paragraph 3.5 for more information.