Keithley 7016A Instruction Manual Download Page 6

 

The instrument and accessories must be used in accordance with its
specifications and operating instructions or the safety of the equip-
ment may be impaired.

Do not exceed the maximum signal levels of the instruments and ac-
cessories, as defined in the specifications and operating informa-
tion, and as shown on the instrument or test fixture panels, or
switching card.

When fuses are used in a product, replace with same type and rating
for continued protection against fire hazard.

Chassis connections must only be used as shield connections for
measuring circuits, NOT as safety earth ground connections.

If you are using a test fixture, keep the lid closed while power is ap-
plied to the device under test. Safe operation requires the use of a
lid interlock.

If a 

 screw is present, connect it to safety earth ground using the

wire recommended in the user documentation.

The 

 symbol on an instrument indicates that the user should re-

fer to the operating instructions located in the manual.

The 

symbol on an instrument shows that it can source or mea-

sure 1000 volts or more, including the combined effect of normal
and common mode voltages. Use standard safety precautions to
avoid personal contact with these voltages.

The 

 

WARNING

 

 heading in a manual explains dangers that might

result in personal injury or death. Always read the associated infor-
mation very carefully before performing the indicated procedure.

The 

 

CAUTION

 

 heading in a manual explains hazards that could

damage the instrument. Such damage may invalidate the warranty.

Instrumentation and accessories shall not be connected to humans.

Before performing any maintenance, disconnect the line cord and
all test cables.

To maintain protection from electric shock and fire, replacement
components in mains circuits, including the power transformer, test
leads, and input jacks, must be purchased from Keithley Instru-
ments. Standard fuses, with applicable national safety approvals,
may be used if the rating and type are the same. Other components
that are not safety related may be purchased from other suppliers as
long as they are equivalent to the original component. (Note that se-
lected parts should be purchased only through Keithley Instruments
to maintain accuracy and functionality of the product.) If you are
unsure about the applicability of a replacement component, call a
Keithley Instruments office for information.

To clean an instrument, use a damp cloth or mild, water based
cleaner. Clean the exterior of the instrument only. Do not apply
cleaner directly to the instrument or allow liquids to enter or spill
on the instrument. Products that consist of a circuit board with no
case or chassis (e.g., data acquisition board for installation into a
computer) should never require cleaning if handled according to in-
structions. If the board becomes contaminated and operation is af-
fected, the board should be returned to the factory for proper
cleaning/servicing.

!

 

Rev. 2/99

Summary of Contents for 7016A

Page 1: ...Instruction Manual Model 7016A 50 2GHz Multiplexer Card Contains Operating and Servicing Information 7016A 901 01 Rev A 5 99...

Page 2: ...OF ITS EMPLOYEES SHALL BE LIABLE FOR ANY DIRECT INDIRECT SPECIAL INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT OF THE USE OF ITS INSTRU MENTS AND SOFTWARE EVEN IF KEITHLEY INSTRUMENTS INC HAS BEEN A...

Page 3: ...Model 7016A 50 2GHz Multiplexer Card Instruction Manual 1999 Keithley Instruments Inc All rights reserved Cleveland Ohio U S A First Printing May 1999 Document Number 7016A 901 01 Rev A...

Page 4: ...orporate immediately into the manual Addenda are num bered sequentially When a new Revision is created all Addenda associated with the previous Revision of the manual are incorporated into the new Rev...

Page 5: ...sent A good safety practice is to expect that hazardous voltage is present in any unknown circuit before measuring Users of this product must be protected from electric shock at all times The responsi...

Page 6: ...y before performing the indicated procedure The CAUTION heading in a manual explains hazards that could damage the instrument Such damage may invalidate the warranty Instrumentation and accessories sh...

Page 7: ...POTENTIAL 6 V CONTACT RESISTANCE 0 5 CONTACT LIFE 3 105 30V 10mA 3 105 900MHz 1W 1 106 cold switching ENVIRONMENT Operating 0 to 50 C up to 35 C at 80 RH Storage 25 C to 65 C EMC Conforms with Europea...

Page 8: ...sistors 1 2 2 Connections and Installation 2 1 Introduction 2 1 2 2 Handling precautions 2 1 2 3 Connections 2 1 2 3 1 Simplified schematic 2 1 2 3 2 Card configuration 2 1 2 3 3 Input output connecti...

Page 9: ...3 1 Environmental conditions 4 2 4 3 2 Recommended verification equipment 4 2 4 3 3 Insertion loss tests 4 2 4 3 4 Isolation tests 4 3 4 3 5 Special handling of static sensitive devices 4 7 4 4 Princi...

Page 10: ...l 7002 2 6 3 Operation Figure 3 1 T connector multiplexer expansion 3 4 Figure 3 2 4 4 matrix connections 3 5 Figure 3 3 Filter testing 3 6 Figure 3 4 Impedance testing 3 7 4 Service Information Figur...

Page 11: ...commended verification equipment 4 2 Table 4 2 Insertion loss values 4 2 Table 4 3 Recommended troubleshooting equipment 4 8 Table 4 4 Troubleshooting procedure 4 8 5 Replaceable Parts Table 5 1 Model...

Page 12: ...zed repair facility in your area for further infor mation When returning the card for repair be sure to fill out and include the service form at the back of this manual in order to provide the repair...

Page 13: ...Keithley part number 7016A 901 00 The manual package includes an instruction manual and any pertinent addenda 1 9 Repacking for shipment Should it become necessary to return the Model 7016A for repair...

Page 14: ...f contamination To avoid this problem operate the card in a clean environment If the card becomes contaminated it should be thoroughly cleaned as explained in paragraph 4 2 2 3 Connections This paragr...

Page 15: ...A Multiplexer A SMA Connector B B Ch 5 Ch 6 B B Ch 7 Ch 8 B Out B Multiplexer B SMB Connector Figure 2 2 Model 7016A configuration Multiplexer B SMB Terminator Connectors Multiplexer B SMA Input Outpu...

Page 16: ...Ch 4 A Out A B B Ch 5 Ch 6 B B Ch 7 Ch 8 B Out B Note All cables are 50 Measuring Instrument AC Signals Figure 2 3 Typical connection scheme 2 3 3 Input output connecting cables All connections to the...

Page 17: ...alled termina tors The terminators recommended in Section 1 are limited to 2W even through the Model 7016A can switch up to 10W Terminator Jack Figure 2 4 50 terminator installation 2 4 Card installat...

Page 18: ...the card all the way forward until the ejector arms engage the mounting cups 4 Push in on the card edge and ejector arms until the card is properly seated 5 Make sure the ejector arms are properly la...

Page 19: ...lation in Model 7002 Ejector Arms locked position CARD 1 CARD 2 CARD 3 CARD 4 INTERCONNECTION INSTALLATION AND REMOVAL OF CARDS BY QUALIFIED SERVICE PERSONNEL ONLY WARNING NO INTERNAL OPERATOR SERVICE...

Page 20: ...ple to close channel 4 of a Model 7016A installed in slot 2 key in the following channel list and press CLOSE SELECT CHANNELS 2 4 To open a closed channel simply press OPEN or OPEN ALL Channels are or...

Reviews: