Keithley 7016A Instruction Manual Download Page 20

 

3

 

Operation

 

3-1

 

3.1

Introduction

 

This section contains basic information on using the Model
7016A including signal limitations and switching consider-
ations. For detailed mainframe operating information, refer
to the Model 7001 or Model 7002 Instruction Manual.

 

3.2

Signal limitations

 

CAUTION

To prevent damage to the Model 7016A,
do not exceed the maximum signal level
specifications of the card. 

 

To prevent over-heating or damage to the relays, never ex-
ceed the following maximum signal levels when using the
Model 7016A:

• Maximum voltage: Any center conductor or shield to

any other center conductor or to chassis: 30V.

• Maximum current: 0.5A per channel.

• Maximum power: 10VA switched up to 900MHz, 3VA

@ 2GHz.

 

3.3

Front panel control

 

3.3.1 Closing and opening channels

 

To close a Model 7016A multiplexer channel, simply key in
the CHANNEL assignment (1-8), then press the Model
7001/7002 CLOSE key. For example, to close channel 4 of a

Model 7016A installed in slot 2, key in the following channel
list, and press CLOSE:

 

SELECT CHANNELS 2!4

 

To open a closed channel, simply press OPEN or OPEN
ALL.

Channels are organized into two multiplexer banks as fol-
lows:

Multiplexer A: channels 1 to 4
Multiplexer B: channels 5 to 8

You can also simultaneously close one channel in each mul-
tiplexer by including both channels in the channel list. For
example, to close channels 1 and 5 of a card in slot 1, enter
the following channel list at the prompt:

 

SELECT CHANNELS 1!1, 1!5

 

(Note that channels are separated by a comma, which can be
inserted by pressing either the ENTER or right cursor key.)

 

NOTE

 

To maintain 50

 

 

 characteristic impedance,

close only one channel per multiplexer si-
multaneously. Closing more than one
channel per multiplexer may cause unex-
pected results. Program the Model 7001/
7002 for restricted operation to avoid clos-
ing more than one channel at a time.

Again, you can open closed channels with the OPEN or
OPEN ALL key. (OPEN opens only channels in the channel
list while OPEN ALL opens all channels.)

Summary of Contents for 7016A

Page 1: ...Instruction Manual Model 7016A 50 2GHz Multiplexer Card Contains Operating and Servicing Information 7016A 901 01 Rev A 5 99...

Page 2: ...OF ITS EMPLOYEES SHALL BE LIABLE FOR ANY DIRECT INDIRECT SPECIAL INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT OF THE USE OF ITS INSTRU MENTS AND SOFTWARE EVEN IF KEITHLEY INSTRUMENTS INC HAS BEEN A...

Page 3: ...Model 7016A 50 2GHz Multiplexer Card Instruction Manual 1999 Keithley Instruments Inc All rights reserved Cleveland Ohio U S A First Printing May 1999 Document Number 7016A 901 01 Rev A...

Page 4: ...orporate immediately into the manual Addenda are num bered sequentially When a new Revision is created all Addenda associated with the previous Revision of the manual are incorporated into the new Rev...

Page 5: ...sent A good safety practice is to expect that hazardous voltage is present in any unknown circuit before measuring Users of this product must be protected from electric shock at all times The responsi...

Page 6: ...y before performing the indicated procedure The CAUTION heading in a manual explains hazards that could damage the instrument Such damage may invalidate the warranty Instrumentation and accessories sh...

Page 7: ...POTENTIAL 6 V CONTACT RESISTANCE 0 5 CONTACT LIFE 3 105 30V 10mA 3 105 900MHz 1W 1 106 cold switching ENVIRONMENT Operating 0 to 50 C up to 35 C at 80 RH Storage 25 C to 65 C EMC Conforms with Europea...

Page 8: ...sistors 1 2 2 Connections and Installation 2 1 Introduction 2 1 2 2 Handling precautions 2 1 2 3 Connections 2 1 2 3 1 Simplified schematic 2 1 2 3 2 Card configuration 2 1 2 3 3 Input output connecti...

Page 9: ...3 1 Environmental conditions 4 2 4 3 2 Recommended verification equipment 4 2 4 3 3 Insertion loss tests 4 2 4 3 4 Isolation tests 4 3 4 3 5 Special handling of static sensitive devices 4 7 4 4 Princi...

Page 10: ...l 7002 2 6 3 Operation Figure 3 1 T connector multiplexer expansion 3 4 Figure 3 2 4 4 matrix connections 3 5 Figure 3 3 Filter testing 3 6 Figure 3 4 Impedance testing 3 7 4 Service Information Figur...

Page 11: ...commended verification equipment 4 2 Table 4 2 Insertion loss values 4 2 Table 4 3 Recommended troubleshooting equipment 4 8 Table 4 4 Troubleshooting procedure 4 8 5 Replaceable Parts Table 5 1 Model...

Page 12: ...zed repair facility in your area for further infor mation When returning the card for repair be sure to fill out and include the service form at the back of this manual in order to provide the repair...

Page 13: ...Keithley part number 7016A 901 00 The manual package includes an instruction manual and any pertinent addenda 1 9 Repacking for shipment Should it become necessary to return the Model 7016A for repair...

Page 14: ...f contamination To avoid this problem operate the card in a clean environment If the card becomes contaminated it should be thoroughly cleaned as explained in paragraph 4 2 2 3 Connections This paragr...

Page 15: ...A Multiplexer A SMA Connector B B Ch 5 Ch 6 B B Ch 7 Ch 8 B Out B Multiplexer B SMB Connector Figure 2 2 Model 7016A configuration Multiplexer B SMB Terminator Connectors Multiplexer B SMA Input Outpu...

Page 16: ...Ch 4 A Out A B B Ch 5 Ch 6 B B Ch 7 Ch 8 B Out B Note All cables are 50 Measuring Instrument AC Signals Figure 2 3 Typical connection scheme 2 3 3 Input output connecting cables All connections to the...

Page 17: ...alled termina tors The terminators recommended in Section 1 are limited to 2W even through the Model 7016A can switch up to 10W Terminator Jack Figure 2 4 50 terminator installation 2 4 Card installat...

Page 18: ...the card all the way forward until the ejector arms engage the mounting cups 4 Push in on the card edge and ejector arms until the card is properly seated 5 Make sure the ejector arms are properly la...

Page 19: ...lation in Model 7002 Ejector Arms locked position CARD 1 CARD 2 CARD 3 CARD 4 INTERCONNECTION INSTALLATION AND REMOVAL OF CARDS BY QUALIFIED SERVICE PERSONNEL ONLY WARNING NO INTERNAL OPERATOR SERVICE...

Page 20: ...ple to close channel 4 of a Model 7016A installed in slot 2 key in the following channel list and press CLOSE SELECT CHANNELS 2 4 To open a closed channel simply press OPEN or OPEN ALL Channels are or...

Reviews: