Service Information
5-2
5.3
Performance verification
The following paragraphs discuss performance verification
procedures for the Model 7015, including path resistance,
offset current, contact potential, and isolation.
With the Model 7015’s backplane jumpers installed, the per-
formance verification procedures must be performed with
only one multiplexer card (the one being checked) installed
in the Model 7001/7002 mainframe. These conditions do not
apply if the backplane jumpers are removed.
CAUTION
Contamination will degrade the perfor-
mance of the card. To avoid contamina-
tion, always grasp the card by the side
edges. Do not touch the connectors, and
do not touch the board surfaces or com-
ponents. On plugs and receptacles, do
not touch areas adjacent to the electrical
contacts.
NOTE
Failure of any performance verification
test may indicate that the multiplexer card
is contaminated. See paragraph 5.2 for in-
formation on cleaning the card.
5.3.1 Environmental conditions
All verification measurements should be made at an ambient
temperature between 18
°
and 28
°
C, and at a relative humid-
ity of less than 70%.
5.3.2 Recommended equipment
Table 5-1 summarizes the equipment necessary for perfor-
mance verification, along with an application for each unit.
Table 5-1
Verification equipment
Description
Model or number
Specifications
Applications
DMM
Keithley Model 2001
2k
Ω
; 50ppm
Channel resistance
Electrometer w/voltage source
Keithley Model 617
10pA, 100pA;
1.6% 100V
source; 0.2%
Offset current, path isolation
Sensitive Digital Voltmeter
Keithley Model 182
3mV; 60ppm
Contact potential
Switching mainframe
Keithley Model 7001 or
7002
All
Triax cable (unterminated)
Keithley Model 7025
Offset current
Low thermal cable (unterminated)
Keithley Model 1484
Contact potential
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for 7015-S
Page 88: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 89: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 91: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 92: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...