Model 6485 Instruction Manual
Calibration
16-13
Displaying calibration dates
To display calibration dates at any time:
1.
From normal display, press MENU, select CAL, then press ENTER. The unit will
display the following:
CAL: VOFFSET
2.
Use either RANGE key to select CAL: DATES, then press ENTER. The Model
6485 will display the last calibration date, for example:
DATE: 11/15/01
3.
Press ENTER to view the calibration due date, for example:
NDUE: 11/15/02
4.
Press EXIT to return to normal display.
Displaying the calibration count
To display the calibration count at any time:
1.
From normal display, press MENU, select CAL, then press ENTER. The unit will
display the following:
CAL: VOFFSET
2.
Use either RANGE key to select CAL:COUNT from the calibration menu, then
press ENTER. The unit displays the calibration count, for example:
COUNT: 1
3.
Press EXIT to return to normal display.
Summary of Contents for 6485
Page 20: ......
Page 24: ......
Page 44: ...1 20 Getting Started Model 6485 Picoammeter Instruction Manual ...
Page 66: ...3 6 Measurements Model 6485 Picoammeter Instruction Manual ...
Page 138: ...9 20 Remote Operation Model 6485 Picoammeter Instruction Manual ...
Page 159: ...11 CommonCommands ...
Page 164: ...11 6 Common Commands Model 6485 Picoammeter Instruction Manual ...
Page 165: ...12 SCPISignalOriented MeasurementCommands ...
Page 190: ...14 12 SCPI Reference Tables Model 6485 Picoammeter Instruction Manual ...
Page 214: ...16 14 Calibration Model 6485 Instruction Manual ...
Page 219: ...A Specifications ...
Page 221: ...B StatusandErrorMessages ...
Page 227: ...C GeneralMeasurement Considerations ...
Page 233: ...D DDCEmulationCommands ...
Page 247: ...E ExamplePrograms ...
Page 250: ...E 4 Example Programs Model 6485 Picoammeter Instruction Manual ...
Page 251: ...F IEEE 488BusOverview ...
Page 266: ...F 16 IEEE 488 Bus Overview Model 6485 Picoammeter Instruction Manual ...
Page 267: ...G IEEE 488andSCPI ConformanceInformation ...
Page 311: ......