Model 6485 and 6487 User’s Manual
Buffer and Sweeps
6-3
Buffer timestamp
To change the buffer timestamp format, press MENU, select TSTAMP, then press ENTER.
Select the desired option: ABS (absolute) or DELT (delta). For ABS, each timestamp is ref-
erenced to the first reading stored in the buffer. The first reading always has a timestamp of
0000000.0000. With DELT, each timestamp provides the time between the readings.
Clear
To clear the Model 6487 buffer, press CONFIG followed by STORE, then set the reading
count to 0 (pressing AUTO will clear the display quickly). Press ENTER with a reading
count of 0 displayed to finish clearing the buffer.
Buffer statistics
•
MIN
and
MAX
provides the minimum and maximum readings stored in the
buffer. It also indicates the buffer location of these readings.
• The
PK-PK (peak-to-peak)
value is the difference between the maximum and
minimum readings stored in the buffer:
PK-PK = MAX - MIN
•
Average
is the mean of the buffer readings. Mean is calculated as follows:
y
n
Σ
X
i
n
-----
i=1
=
Where: X
i
is a stored reading.
n is the number of stored readings.
• The
STD DEV
value is the standard deviation of the buffered readings. Standard
deviation is calculated as follows:
y
n
Σ
Avg X
i
–
(
)
2
i 1
=
n 1
–
------------------------------------
=
Where:
X
i
is a stored reading.
n
is the number of stored readings.
Avg
is the mean of the buffer readings.
Summary of Contents for 6485
Page 40: ...1 28 Getting Started Model 6485 and 6487 User s Manual...
Page 94: ...7 8 Remote Operation and Commands Model 6485 and 6487 User s Manual...
Page 95: ...A Specifications...
Page 98: ...A 4 Specifications Model 6485 and 6487 User s Manual...
Page 105: ...C ExamplePrograms Programming examples lists and defines three types of programming exam ples...
Page 113: ......
Page 114: ......