![Keithley 6485 User Manual Download Page 64](http://html2.mh-extra.com/html/keithley/6485/6485_user-manual_4060113064.webp)
3-8
Measurements and Sourcing Voltage
Model 6485 and 6487 User’s Manual
Figure 3-2
Connections for Model 6487 current measurements
Black (LO)
Red (HI)
237-ALG-2
Cable
Metal Noise Shield
Metal Safety Shield
Safety
Earth
Ground
Green
(Chassis)
Model 6487
120
LINE RATING
50, 60Hz
50 VA MAX
INPUT
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
IEEE-488
CAT
I
TRIGGER LINK
RS-232
MADE IN
U.S.A.
ANALOG OUT
DIGITAL I/O
INTERLOCK
505V
MAX
V-SOURCE OUTPUT
505V PK TO CHASSIS
LO
HI
!
!
FUSE
LINE
630mAT
(SB)
100 VAC
120 VAC
315mAT
(SB)
220 VAC
240 VAC
505V PK
505V PK
505V PK
SCPI programming
Table 3-4
SCPI commands — basic current measurements
Commands*
Description
FUNC ‘CURR’
RANG <Range>
RANG:AUTO <b>
INIT
READ?
Select current function (Model 6487 only).
Select manual range (-0.021 to 0.021A).
Enable (ON) or disable (OFF) auto range.
Trigger reading.
Trigger and return reading(s).
* Zero correct and zero check commands not included. See
— current measurements
Summary of Contents for 6485
Page 40: ...1 28 Getting Started Model 6485 and 6487 User s Manual...
Page 94: ...7 8 Remote Operation and Commands Model 6485 and 6487 User s Manual...
Page 95: ...A Specifications...
Page 98: ...A 4 Specifications Model 6485 and 6487 User s Manual...
Page 105: ...C ExamplePrograms Programming examples lists and defines three types of programming exam ples...
Page 113: ......