19-6
Calibration
Mainframe calibration
Use the procedures discussed below to calibration the mainframe without the Remote
PreAmp. See
Remote PreAmp calibration
later in this section for information in calibrating the
Remote PreAmp.
NOTE
The mainframe must be separately calibrated before calibrating the Remote
PreAmp.
Mainframe calibration menu
You can access the calibration menu by pressing the MENU key and then selecting CAL in
the main menu. Mainframe calibration menu selections include:
•
UNLOCK: Enables calibration.
•
EXECUTE: Allows you to execute the calibration procedure for the selected function
and range.
•
VIEW-DATES: Displays calibration and calibration due dates.
•
SAVE: Saves new calibration constants.
•
LOCK: Disables calibration.
•
CHANGE-PASSWORD: Allows you to change the calibration password.
Mainframe calibration procedure
The mainframe calibration procedure described below calibrates all ranges of both the cur-
rent and voltage source and measure functions. (Resistance calibration is not required.) Note
that each range is separately calibrated by repeating the entire procedure for each range.
Step 1: Prepare the Model 6430 for calibration
1.
With the power off, disconnect the Remote PreAmp from the mainframe.
2.
Turn on the Model 6430 and the digital multimeter, and allow them to warm up for at
least one hour before performing calibration.
3.
Press the MENU key, then choose CAL and press ENTER. Select UNLOCK, and then
press ENTER. The instrument will display the following:
PASSWORD:
Use
,
▲
,
▼
,
, ENTER or EXIT.
4.
Use the up and down range keys to select the letter or number, and use the left and right
arrow keys to choose the position. Enter the present password on the display. (Front
panel default: 006430.) Press ENTER to complete the process.
5.
Press EXIT to return to normal display. Instrument operating states will be set as sum-
marized in Table 19-2.
▲
▲
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......