background image

2-3. 

VOLTAGE  MEASUREMENTS. 

B. 

Normal 

Measurements.  When  the  RANGE 

switch  is  set  to  VOLTS  position  the  610e  can  be  used 
to  measure  voltages  from  .001  volt  full  scale  to  100 

volts.  In  this  mode  the  input  impedance  is  greater 
than  1014  ohms  shunted  by  20  picofarads.  The  input 

impedance  can  be  reduced  when  necessary  by  setting  the 

RANGE  switch  to  AMPERES  as  discussed  in  paragraph  2-3c. 

1.  Set  the  controls  as  follows: 

ZERO  CHECK  Switch 

RANGE  Switch 

Multiplier  Switch 

FEEDBACK  Switch 
METER  Swi t ch 

LOCK 
VOLTS 

100 

NORMAL 
CENTER  ZERO 

2. 

Connect  the  input  source  to  the 

UHF 

input. 

Unlock  the  ZERO  CHECK  switch  to  make  a  reading. 

Increase  the  meter  sensitivity  using  the  Multiplier 
switch  to  obtain  a  useable  on-scale  reading. 

The 

METER  switch  should  be  set  to  either 

or  - to 

obtain  the  rated  voltage  accuracy. 

b.  Guarded 

Measurements.  This  method  is 

useable  when  a  Keithley  Model  6108  input  cable  is 
available. 

The  Model  6108  is  a  specialized  triaxial 

cable  with  a  mating 

UHF 

connector.  The  cable  is 

terminated  with  three  color-coded  alligator  clip  leads. 

The  leads  are  identified  as  shown  in  Table  4. 

TABLE  4. 

Model  6108  Input  Cable. 

Lead  Color 

Terminal  Function 

Red 

Input  High 

Black 

Inner  Shield  (Guard) 

Green 

Outer  Shield  (Ground) 

1. 

Set  the  controls  as  follows: 

ZERO  CHECK  Switch 

RANGE  $wi tch 

Multiplier  Switch 
FEEDBACK  Switch 

METER  Swi tch 

LOCK 
VOLTS 
100 
NORMAL 
CENTER  ZERO 

2. 

Connect  the  Model  6108  to  the  UHF  input. 

Connect  the  black  banana  plug  to  the  Xl  OUTPUT 

on  the  rear  panel  of  the  instrument.  Connect  the 

remaining  alligator  clip  leads  as  shown  in  Figure 

5. 

c.  Decreased 

Resistance. 

The  input  resistance 

can  be  reduced  for  applications  where  very  high  input 

resistance  is  not  required.  Set  the  RANGE  switch  to 

one  of  the  AMPERES  ranges  where  the  input  is  shunted 

by  a  resistor. 

The  input  resistance  is  then  the  re­

ciprocal  value  of  the  AMPERES  setting.  For  instance, 
to  obtain  an  input  resistance  of  10  megohms  (107n) 

set  the  RANGE  switch  to  the  10-7  AMPERES  range.  Then 
adjust  the  Multiplier  switch  to  the  full  scale  volt­
age  desired.  The  6l0C  can  now  be  used  for  voltage 
measurements  as  discussed  in  paragraphs  2-3a  and  b. 

d.  High  Voltage  Measurements.  Measurements  can 

be  made  up  to 

30 

kilovolts  using  one of  the  Voltage 

Divider  probes  listed  in  Table  5.  The  full  scale 
voltage  range  is  determined  by  multiplying  the  divider 

ratio  times  the  Multiplier  s"etting. 

TABLE  S 

Voltage  Divider  Probes 

Model 

Divider  Ratio 

Accuracy 

6102A 

6103A 

5%  at  10  kv 

2-4.  CURRENT  MEASUREMENTS. 

a.  Normal  Current  Measurements.  When  the  FEEDBACK 

switch  is  set  to  NORMAL  the  current  is  determined .by 
measuring  the  voltage  drop  across  a  RANGE  resistor 
shunting  the  6l0C  input.  This  method  is  useable  for 
any  current  range  where  low  noise  is  essential. 

1.  Set  the  front  panel  controls  as  .follows: 

ZERO  CHECK  Switch 
RANGE  Swi tch 
Multiplier  Switch 

FEEDBACK  Switch 
METER  Swi t ch 

LOCK 
10-1  AMPERES 

NORMAL 

CENTER  ZERO 

2.  Connect  the  610C  input  terminals  in  series  with 

the  current  to  be  measured.  Unlock  the  ZERO  CHECK 
switch  to  make  a  reading.  Increase  the  sensitivity 
using  the  RANGE  switch.  The  full  scale  current  is 
determined  by  multiplying  the  RANGE  setting  times 
the  Multiplier  setting.  Use  the  smallest  Multiplier 
setting  to  obtain  the  best  accuracy:  The  input  voltage 

drop  of  the  610C  in  the  NORMAL  mode  is  a  function  of 

the  Multiplier  setting  and  therefore  the  loading 
error  should  be  noted  as  shown  in  Figure  6. 

TABLE  6. 

Response  Speed  For  Fast  Current  Measurements. 

Range 

Rise  Time 

Noise  (Peak-to-Peak) 

� 

No  C 

10-l4to 

2.0 

2.0 

2.0 

10-15 

loS 

10-14 

10-13 

10-11 
10-10 

.45 

.4S 

.4S 

8.5 

10-14 

10-13 

2.5 

10-12 

10-9 

.07 

.07 

.07 

6.8 

10-13 

3.2 

10-12 

10-11 

10-8 

.01S 

.01S 

.01S 

10-12 

2.S 

10-11 

10-10 

10-7 

.003 

.003 

.003 

10-11 

10-11 

10-10 

10-6 

.003 

.003 

.003 

loS 

10-10 

3.3 

10-10 

8.S 

10-10 

Summary of Contents for 610C

Page 1: ...Model 610C 610CRSolid State Electrometers Instruction Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...

Page 2: ...RECT SPECIAL INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT OF THE USE OF ITS INSTRUMENTS AND SOFTWARE EVEN IF KEITHLEY INSTRUMENTS INC HAS BEEN ADVISED IN ADVANCE OF THE POSSIBILITY OF SUCH DAMAGES...

Page 3: ...ates that a shock hazard exists when voltage levels greater than 30V RMS 42 4V peak or 60VDC are present A good safety practice is to expect that hazardous voltage is present in any unknown circuit be...

Page 4: ...components in mains circuits including the power transformer test leads and input jacks must be purchased from Keithley Instru ments Standard fuses with applicable national safety approvals may be use...

Page 5: ...Model 610C 610CR Solid State Electrometers Instruction Manual Keithley Instruments Inc All rights reserved Cleveland Ohio U S A Third Printing December 2001 Document Number 29113 Rev C...

Page 6: ...i CONTENTS Section Page SPECIFICATIONS ii 1 GENERAL DESCRIPTION 1 2 OPERATION 4 3 CIRCUIT DESCRIPTION 9 4 ACCESSORIES 12 5 SERVICING 15 6 CALIBRATION 24 7 REPLACEABLE PARTS 29 SCHEMATIC 37 13 22 27 35...

Page 7: ...ric Measurements Resistance Measurements Recorder Output 100 mV Divider Network Voltmeter Configuration NORMAL Mode Ammeter Configuration FAST Mode Ohmmeter Configuration NORMAL Mode Chassis Side View...

Page 8: ...opposite input polarity GAIN 0 1 0 3 etc to 3000 FREQUENCY RESPONSE within 3db DC to 40kHz at a gain of I and lower decreasing to DC to 100Hz at maximum gain Full output response limited to 3kHz on an...

Page 9: ...ess than 1 millivolt per 24 hours Zero offset due to temperature change is less than 150 microvolts per C after a 30 minute warm up This offset how ever can easily be compensated for with the front pa...

Page 10: ...used with AMPERES OHMS or COULOMBS the setting should be multiplied by the RANGE setting Controls power to instrument Also selects the meter polarity and center scale Meter zero adjustment Meter zero...

Page 11: ...EDIUM 5101 SWITCH S107 RANGE SWITCH 5106 ZERO CHECK SWITCH S103 ZERO CONTROL POWER COARSE S102 NEON FEEDBACK SWITCH 5105 GROUND J102 INPUT J101 DS101 FIGURE 3 Front Panel Controls Model 6l0eR 1MA CAL...

Page 12: ...he FEEDBACK Switch at FAST some insulators such as Teflon may produce ran 10m signals which show up as erratic meter deflections nsulation used in the Model 6l0C is carefully select zd to minimize the...

Page 13: ...e re ciprocal value of the AMPERES setting For instance to obtain an input resistance of 10 megohms 107n set the RANGE switch to the 10 7 AMPERES range Then adjust the Multiplier switch to the full sc...

Page 14: ...CABLE CONNECT UNKNOWN BETWEEN HIGH AND GND r I I I FIGURE 5 Guarded Voltage Measurement T IS ES RS ERROR IS 1M ERROR EIN ES x 100 IF EIN ES THEN 1 M IS FIGURE 6 Current Measurements Loading Error 1 V...

Page 15: ...be The circuit connections are shown in Figure 8 mine the resistance the Multiplier setting subtracted from the voltage source setting resistance is therefore the ratio of the net divided by the curre...

Page 16: ...r MA CAL MODEL 610C I I I I I I I I J NOTE In the FAST mode the OUTPUT is floating with respect to chassis ground 2 Current Output 1 milliampere for full scale input variable 5 with 1400 ohm recorder...

Page 17: ...of the Model 6l0C is as follows UHF INPUT 1 The amplifier is always in a unity gain input voltage to output current converter configuration The internal circuitry is arranged such that a fu11 scale in...

Page 18: ...f input capacity it also reduces the input drop to less than 100 microvolts 10 3 5 OIlMl ETER OPERATION a Normal Method In the normal method of resis tance measurements FEEDBACK Switch in NORMAL posi...

Page 19: ...UHF INPUT 11 A OUTPUT RANGE MULTIPLIER GUARD FIGURE 11 Ammeter Configuration FAST Mode xl OUTPUT UHF INPUT OUTPUT MULTIPLIER FIGURE 12 Ohmmeter Configuration NORMAL Mode II...

Page 20: ...ading Model 6105 Resistivity Chamber The Model 6105 is a guarded test fixture for measurement of surface and volume resistivity The chamber is designed in accordance with ASTM Standard Method of Test...

Page 21: ...ace components only as necessary Use only reli able replacements which meet the specifications b The MOS FET input transistors QIOI and QI02 are specially selected and matched order only as a plug in...

Page 22: ...h at VOLTS Multiplier Switch at 1 and the meter zeroed Measurements are with the Model 153 c At times the meter will not zero on any range with the input shorted Adjust the COARSE ZERO Switch Sl02 Fig...

Page 23: ...nd adjusting the COARSE and MEDIUM ZERO Controls for null If null is not reached check this stage and check for shorts in the base circuit of Ql07 and Ql08 5 Connect Model 153 to the collector of Q109...

Page 24: ...FIGURE 14 Chassis Side View 18 16...

Page 25: ...FIGURE 15 Circuit Board pelS3 19 17...

Page 26: ...FIGURE 16 Printed Circuit Board PC 186 20 18...

Page 27: ...rr rr 6 0 N a IT IT ct 610C SWITCH 5101 o N rf rf rf r a a FIGURE 17 Switch MEDIUM ZERO S101 ct a ct 19 N IT I D 6 N N N Ii r1 r1 610C SWITCH 5102 FIGURE 18 Switch COARSE ZERO 5102 DECK 2 DECK DECK 2...

Page 28: ...IO 4 jl We h R IOS I 0 jl aiC h I R IOG jL h dCI07 I RI07 dCI06 6iC h I RIOS UiC H RIO 9 f i C h RIIO J I L f 1 Rill I b 0 NC h R 1I2 i J c P P ocp IJ I C 10Z I I CI03 I j db I IJ CI04 6 db IJ C lOS I...

Page 29: ...0 CJ CJ 0 Tl m rn ITI n n A A A N VJ 0 0 RI75 I I l RI7 Jl c u R17 3 i R t72 c jl Jl 0 J RI70 RlioS CI23 C 2 o R IGG C 1 2 1 FIGURE 20 Switch Multiplier Sl07 23 21...

Page 30: ...CALIBRATION Zero whenever other adjustments are made a To set the mechanical Zero meter turn the METER Switch to METER OFF and set the mechanical Zero Meter adjustment for zero meter reading top scale...

Page 31: ...rent of the Model 6l0C set the front panel controls to ZERO CRECK Button Multiplier Switch Range Switch FEEDBACK Switch METER Switch LOCK 003 10 11 AMPERES FAST Cap the INPUT Receptacle and unlock the...

Page 32: ...from 10 6 to 10 8 am ere ranges and 4 of full scale from the 10 9 to 10 1 ampere ranges For the ranges above 10 5 ampere construct a current source whose output is 0 2S Set the Model 610C FEEDBACK Swi...

Page 33: ...t range 1 Set the Signal Generator to 20 Hz at minimum output Release the Electrometer ZERO CHECK Button and increase the Generator output to obtain 2 volt rms at the Model GlOC output 2 Set the Gener...

Page 34: ...26...

Page 35: ...ETB F Fig GCb k M or meg Mfg Circuit DesiS C10l C102 C103 C104 C10S C106 C107 C108 C109 CllO Clll Cll2 Cll3 Abbreviations and Symbols ampere Ceramic disc Composition Deposited Carbon Electrolytic Alum...

Page 36: ...Pilot Light Mfg No 53 Pilot Light Holder Mfg No 203CE Fuse slow blow 1 4A 3AG Mfg No 313 250 Fuse slow blow 1 8A 3AG Mfg Type MDL Fuse holder Mfg No 342012 Receptacle UHF INPUT Mfg No 6804 F P1ug UHF...

Page 37: ...t Assembly Zero Check Button 80164 80164 KN 47 16988A 1 3 14 S104 Slide Switch 117 234V 80164 SW 151 2 4 S105 Slide Switch FEEDBACK 80164 SW 45 1 3 14 S106 Rotary Switch less components RANGE Rotary S...

Page 38: ...Cb 07716 DCC RI2 68K 17 R126 249 n 1 1 2 W DCb 91637 DCF 1 2 249 Rl2 249 18 R127 249 n 1 1 2 W DCb 91637 DCF 1 2 249 R12 249 18 R128 249 n 17o 1 2 W DCb 91637 DCF 1 2 249 R12 249 18 R129 249 n 1 1 2 W...

Page 39: ...WI 01686 7009 R95 9 1 20 R174 2 73 n 1 1 4 I WI 01686 7009 R95 2 73 20 R175 91 Q 1 1 4 I WI 01686 7009 R95 91 20 R176 8 6 kn 1 1 2 I DCb 07716 ncc R12 S 6K 14 RI77 20 kn 10 1 5W WllVar 71450 AI RP34 2...

Page 40: ...l I 17076B Handle Assembly I HH 18 2 Screw t 1 8 32xl 4 slotted 4 D8 32xl 4 slotted P li 3 Bottom Cover Assembly I 21540C Bottom Cover Sheet Metal I 2154lC Tilt Bail I 15004B Support Assembly Right I...

Page 41: ..._ v 3 FIGURE 21 Chassis Assembly 0 1r 11 35 33...

Page 42: ...N Wesson Miss 39191 44655 Ohmite Mfg Co 3601 Howard St Skokie Ill 60476 56289 Sprague Electric Co North Adams Mass 58474 Superior Electric Co The 383 Middle St Bristol Conn 06012 63060 Victoreer 0 tr...

Page 43: ......

Page 44: ...Landsberger Strasse 65 D 82110 Germering 089 84 93 07 40 Fax 089 84 93 07 34 GREAT BRITAIN Keithley Instruments Ltd The Minster 58 Portman Road Reading Berkshire RG30 1EA 0118 9 57 56 66 Fax 0118 9 5...

Reviews: