4. Press the navigation wheel
to enter EDIT mode. The EDIT
indicator appears in the upper left corner of the display.
5. Turn the navigation wheel
to set the source value to
100.0000 V
, and then press the navigation wheel
to enter the
selection and exit EDIT mode.
Step 2: Set the compliance limit
1. Press the
LIMIT
key. You will see a blinking character in
the LimA value field.
2. Press the UP range arrow once so that the 10s digit is displayed.
It is displayed as 01.0000mA.
3. Use the cursor key to highlight the 1s digit.
4. Press the navigation wheel
to enter EDIT mode.
5. Use the navigation wheel to set the limit to 10 mA.
6. Press the navigation wheel again or press ENTER to enter the
selected value and exit EDIT mode.
Step 3: Set measurement function and range
1. Press the
MEAS
key as many times as needed to select the
V
(voltage) measurement function. In the following figure, the
measurement function has been set to volts.