Wiring the interlock  

The Model 2657A is a high voltage instrument that can output 
hazardous live voltages. It is designed to be used with a test fixture or 

in a system in which operators are not exposed to such hazardous 

voltages. 

The Model 2657A provides an interlock as a method for ensuring 

operator safety. The output of the Model 2657A can only be turned 

on when the interlock line is enabled. If you attempt to turn on the 
output when the interlock is disabled, the instrument will generate 

error code 802, “Output blocked by interlock.” 

The interlock line is enabled when it is pulled high through a switch 

to > +4 V. To drive the interlock pin high, the external supply must 

supply a minimum of 50 mA. The interlock line is disabled when the 
signal applied is < +4 V. The absolute maximum input is 

0.4 V to +6.0 V.  

The interlock is intended for use through a normally open switch. 
This switch may be installed on the lid of a test fixture, the enclosure 

of a semiconductor prober or device handler, or on the doors of a 

test equipment rack. 

 

The interlock line is pin 24 of the digital input/output connector on 

the rear panel of the Model 2657A. 

 

 

A +5.0V pin is also available on the digital I/O connector (pin 22). 
To use the interlock, connect the 5 V pin and the interlock pin to the 

input and output respectively of the switch in your test system. 

The following graphic shows how the interlock is intended to be used 
with a generic test fixture. 

If you are using the Model 2657A with the Model 8010 High Power 
Device Test Fixture, the interlock connection may be made using the 
CA-558-2 cable assembly. Refer to the Model 8010 User's Manual for 
information on how to connect the Model 2657A to the Model 8010. 

 
 

Summary of Contents for 2657A

Page 1: ...Model 2657A High Power System SourceMeter Instrument Quick Start Guide...

Page 2: ...work on live circuits perform safe installations and repair products Only properly trained service personnel may perform installation and service procedures Keithley Instruments products are designed...

Page 3: ...in the specifications and operating information and as shown on the instrument or test fixture panels or switching card When fuses are used in a product replace with the same type and rating for cont...

Page 4: ...not apply cleaner directly to the instrument or allow liquids to enter or spill on the instrument Products that consist of a circuit board with no case or chassis e g a data acquisition board for inst...

Page 5: ...ion about the instrument s features and programming commands The User s and Reference manuals are in PDF format on the CD ROM that is included with the instrument If you do not have Adobe Reader to vi...

Page 6: ...Unpack and inspect the instrument To unpack and inspect the instrument 1 Inspect the box for damage 2 Open the top of the box...

Page 7: ...emove the packaging insert 4 Carefully lift the Model 2657A instrument out of the box 4 Inspect the instrument for any obvious signs of physical damage Report any damage to the shipping agent immediat...

Page 8: ...low ground cable CA 568 120 4 RJ45 LAN crossover cable 5 Interlock DB 25 male connector kit hardware 6 Model 2657A Quick Start Guide not shown this document 7 Model 2657A Product Information and Test...

Page 9: ...only rated for dry locations Read and comply with the specifications of all instruments in the system The overall allowed signal levels may be constrained by the lowest rated instrument in the system...

Page 10: ...eep users safe always read and follow all safety warnings provided with each of the instruments in your system Install the instrument The Model 2657A can be used on a bench or in a rack Please see the...

Page 11: ...applied is 4 V The absolute maximum input is 0 4 V to 6 0 V The interlock is intended for use through a normally open switch This switch may be installed on the lid of a test fixture the enclosure of...

Page 12: ...rth safety ground Nonconductive test fixtures must be rated to double the maximum voltage in the test circuit Failure to attach the ground wires to a known protective earth may result in electric shoc...

Page 13: ...hassis is connected to power line ground through the ground wire in the power cord Failure to use a grounded outlet may result in personal injury or death due to electric shock Operating the instrumen...

Page 14: ...nce When the instrument is turned on you should see Five dots All segments of the display light A brief display showing KEITHLEY Model 2657A Other startup checks The entire power up process takes appr...

Page 15: ...d You do not need to connect a device under test DUT for this test Step 1 Set source function range and level 1 Press the SRC key You will see a blinking character in the SrcA value field Confirm that...

Page 16: ...A value field 2 Press the UP range arrow once so that the 10s digit is displayed It is displayed as 01 0000mA 3 Use the cursor key to highlight the 1s digit 4 Press the navigation wheel to enter EDIT...

Page 17: ...ements begin Step 5 Observe measurements Observe the measured voltage on the main area of front panel display The readings should be very close to the 100 V source value Step 6 Turn output off When yo...

Page 18: ......

Page 19: ...priate to the final shipping destination of the instrument This line frequency is used for aperture NPLC calculations for the integrating analog to digital converter ADC You can manually configure the...

Page 20: ...or to a safe high voltage SHV connector What should I do if I get an 802 interlock error You will receive error code 802 Output blocked by interlock if You disable the interlock when the Model 2657A o...

Page 21: ...the Product Information CD ROM that came with your instrument Model 2657A Reference Manual Refer to the Model 2657A Reference Manual for detailed information about all features of the instrument This...

Page 22: ......

Reviews: