Model 2651A High Power System SourceMeter® Instrument Reference Manual
Section 7: Command reference
2651A-901-01 Rev. A / March 2011
7-197
Details
This attribute sets the size of the stack used for filtered measurements.
Example
smua.measure.filter.count = 10
Sets the filter count for SMU Channel A to 10.
Also see
(on page 7-197)
(on page 7-198)
(on page 7-209)
smuX.measure.filter.enable
This attribute enables or disables filtered measurements.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
SMU reset
Recall
Saved setups
smuX.FILTER_OFF
Usage
filterState
= smu
X
.measure.filter.enable
smu
X
.measure.filter.enable =
filterState
filterState
The filter status. Set to one of the following values:
0
or
smu
X
.FILTER_OFF
: Disables the filter
1
or
smu
X
.FILTER_ON
: Enables the filter
X
SMU channel (for example,
smua.measure.filter.enable
applies to SMU
Channel A).
Details
This attribute enables or disables the filter.
Example
smua.measure.filter.enable = 1
Enables the filter for SMU Channel A. Alternatively, the
value 1 may be replaced with
smua.FILTER_ON
.
Also see
(on page 7-198)
(on page 7-209)