3
Selecting the operating (display) mode:
Actual V and I Mode:
This mode sets the instrument to display steady-state readback current
and voltage and to return current or voltage measurements over the GPIB bus. If
using the front panel controls, press the DISPLAY button and select the ”Actual V
and I” mode using the
and
buttons.
GPIB1,2 commands for selecting the Actual V and I mode:
:sense:function “voltage”
; selects voltage
measurement function
:sense:average xx
; the number of readings to
average per measurement, up
to 10 (optional command)
or
:sense:function “current”
; selects current
measurement function
:sense:average xx
; the number of readings to
average per measurement, up
to 10 (optional command)
Figure 2
. Rear panel of the Model 2304. Note the location of the digital voltmeter
inputs.
5
H I G H S P E E D P R E C I S I O N R E A D B A C K P O W E R S U P P L Y
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
LINE RATING
100-240VAC
50, 60 HZ
185VA MAX
LINE FUSE
SLOWBLOW
2.5A, 250V
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
REMOTE
DISPLAY
OPTION
ISOLATION FROM EARTH:
22 VOLTS MAX.
+
+
+
_
_
_
_
+
SOURCE
SENSE
DVM
IN
SOURCE
OUTPUT
0-20V, 0-5A