:sense:pcurrent:mode low
; set up Model 2304 to
calculate and return low
pulse value
or
:sense:pcurrent:mode average ; set up Model 2304 to
calculate and return
average pulse value
:sense:pcurrent:synchronize: delay .xxxxx
; specify trigger delay in
sec, 0.1 sec maximum in 10
µs increments (optional
command)
GPIB commands to digitize a current waveform and transfer the readings via the bus:
:sense:function “pcurrent”
; selects pulsed current
measurement function
:sense:pcurrent:synchronize:
tlevel x.xxxx
; specify trigger level, up
to 5 amps
:sense:pcurrent:synchronize:
state off
; turn the trigger
synchronization off
:sense:pcurrent:average xxxx ; number of readings to
store in the buffer, up to
5000
:output:response enhanced
; select enhanced output
for pulsed current
measurements
:output on
; turn on supply output
:read:array?
; command used to return
the last array of readings
Note:
This command must be used in place of the :read? command.
Conversion rate is
≈
3600 rdgs/s.
1 1
H I G H S P E E D P R E C I S I O N R E A D B A C K P O W E R S U P P L Y