GPIB commands for making pulse-current measurements:
:sense:function “pcurrent”
; selects pulsed current
measurement function
:sense:pcurrent:synchronize: tlevel x.xxxx
; specify trigger level
x.xxxx, up to 5 amps
:sense:pcurrent:average x
; number of readings to
average per measurement, up
to 100 (optional command)
:output:response enhanced
; select enhanced output
for pulsed current
measurements
:output on
; turn on supply output
and
:sense:pcurrent:time:high .xxxxxx
; specify integration time
(in sec) for high pulse
measurements
:sense:pcurrent:time: low .xxxxxx
; specify integration time
(in sec) for low pulse
measurements
:sense:pcurrent:time: high .xxxxxx
; specify integration time
(in sec) for average pulse
measurements
or
:sense:pcurrent:time:auto
; Model 2304 sets
integration times
and
:sense:pcurrent:mode high
; set up Model 2304 to
calculate and return high
pulse value
or
M O D E L 2 3 0 4