Getting Started 1-9
NOTES
“ON” indicates that the output is turned on. With the output turned off, “OFF” is dis-
played. See Section 10 for details.
For the Pulse Current and Long Integration display modes, “NO PULSE” is
displayed if the output is off or pulses are not detected (output on). See Sections 3 and
4 for details.
When a change is made that affects the readings being taken, dashes are displayed
instead of readings. The dashes remain until a valid reading for the new condition is
taken.
Any one of the four display modes can be the power-on default. Use the SAVE SETUP
item of the MENU to save the selected display mode in memory, and use the POWER
ON SETUP item to specify the power-on setup (see “Setups” under “Default set-
tings” for details).
Default settings
The power supply can be set to power-on to the factory default conditions (RST defaults) or
to user-saved setup conditions. The factory default conditions are listed in Table 1-1.
Table 1-1
Factory defaults (RST)
Setting
RST default
Output value settings:
Voltage (V)
0.000V
Current (A)
0.2500A
Output state operate
Off
Display type
Actual V and I
GPIB address
No effect (factory set to 16)
GPIB output format
No effect (factory set to Keithley
and Exponential)
Current range
5 amps (Auto Range off)
Integration rate
1.00 PLC
Average readings
1
Power on setup
No effect (factory set to RST)
Current limit mode
Lim
Output relay
No effect (after power cycle, set
to zero)
Pulse current:
High time
33 µsec
Low time
33 µsec
Summary of Contents for 2303
Page 28: ...1 14 Getting Started...
Page 44: ...2 16 Basic Power Supply Operation...
Page 64: ...4 10 Long Integration Measurements...
Page 70: ...5 6 Relay Control...
Page 102: ...7 20 Status Structure...
Page 103: ...8 CommonCommands...
Page 108: ...8 6 Common Commands...
Page 109: ...9 Signal Oriented Measurement Commands...
Page 121: ...11 SCPITables...
Page 129: ...A Specifications...
Page 132: ...A 4 Specifications...
Page 133: ...B ErrorandStatusMessages...
Page 137: ...C EmulationCommands...
Page 143: ...D Models2303and2303B GPIB488 1Protocol...
Page 151: ......