7-12
Status Structure
Operation event status
The used bits of the operation event register (shown in Figure 7-5) are described as follows:
•
Bit B3, current limit (CL)
— Set bit indicates that the output is in current limit. This
bit clears when the instrument is no longer in current limit.
•
Bit B4, current limit tripped (CLT)
— Set bit indicates that the output has turned off
due to a current limit condition. This bit clears when the output is turned back on.
•
Bit B6, power supply shutdown (PSS)
— This bit indicates that the output has turned
off due to an overload and/or overheat condition. This typically indicates that the user
tried to exceed the power output capability of the unit.
___
PSS
___
CLT
CL
___
(B15-B7)
(B6)
(B5)
(B4)
(B3)
(B2-B0)
OR
&
&
&
:CONDition?
[:EVENt]?
To OPC bit
of Status Byte
Register
:ENABle <NRf>
:ENABle?
Decimal Weights
Operation Condition
Register
Operation Event
Register
PSS = Power Supply Shutdown
CLT = Current Limit Tripped
& = Logical AND
OR = Logical OR
___
PSS
___
CLT
CL
___
(B15-B7)
(B6)
(B5)
(B4)
(B3)
(B2-B0)
___
PSS
___
CLT
CL
___
(B15-B7)
(B6)
(B5)
(B4)
(B3)
(B2-B0)
___
64
___
16
8
___
(2
6
)
(2
4
)
(2
3
)
Operation Event
Enable Register
CL = Current Limit
Figure 7-5
Operation event status
Summary of Contents for 2303
Page 28: ...1 14 Getting Started...
Page 44: ...2 16 Basic Power Supply Operation...
Page 64: ...4 10 Long Integration Measurements...
Page 70: ...5 6 Relay Control...
Page 102: ...7 20 Status Structure...
Page 103: ...8 CommonCommands...
Page 108: ...8 6 Common Commands...
Page 109: ...9 Signal Oriented Measurement Commands...
Page 121: ...11 SCPITables...
Page 129: ...A Specifications...
Page 132: ...A 4 Specifications...
Page 133: ...B ErrorandStatusMessages...
Page 137: ...C EmulationCommands...
Page 143: ...D Models2303and2303B GPIB488 1Protocol...
Page 151: ......