Model 2290-5 5 kV Power Supply User's Manual
Section 2: Safe configuration and test setup
2290-5-900-01 Rev. A/December 2013
2-3
Connecting the interlock of the Model 2290-5
The interlock pins control the high-voltage supply.
The pin locations and connections are shown in the following figure. The pins are:
•
Pin 3: Chassis ground
•
Pin 2: Interlock enable
•
Pin 1: +5 V DC out (current limited)
Figure 3: Model 2290-5 Interlock connector wiring
The following figure shows how to connect the Model 2290-5 internal 5 V source in order to engage
the external interlock (or interlocks) circuits:
Figure 4: Model 2290-5 Internal 5 kV source
Connecting the Model 2290-5 output to a custom test fixture
Keithley Instruments offers several accessories that can help you when building a custom test fixture
or system.
You can use a high-voltage SHV bulkhead connector (Model 2290-5-SHVBH) on your test fixture.
The following sections show you how to connect a Model 2290-5 to a custom fixture and also how to
make connections to the device inside the custom fixture.
Using SHV coaxial connections
If you need to convert from safe high-voltage (SHV) to another connector type, be sure to
only adapt SHV connectors to connectors that are rated to the maximum possible voltage in
your test setup. If you use adapters that are not rated to the maximum possible voltage in
your test setup, electric shock may result.