Card Connections and Installation
2-13
Current connections
shows typical connections for measuring currents using the
Model 2000-SCAN-20 with current shunts installed. As noted previously, only channels 1 and
11 have current shunts installed. This configuration can be used to scan the following:
•
DC current
•
AC current
Figure 2-8
Current connections
HI
LO
HI
LO
HI
LO
HI
LO
HI
LO
HI
LO
HI
LO
CH 1
CH 2
CH 10
CH 11
CH 12
CH 20
OUT A
To DMM Input
2000-SCAN-20 Card
AC or DC
Current
HI
LO
HI
LO
OUT B
J1036
J1035
A
A
AC or DC
Current
NOTE: Current shunts
must be installed
for channels 1
and 11.