manualshive.com logo in svg
background image

 

 

Halo Photonics 

Stream Line

XR

 Scanning Doppler LiDAR system - Hardware Guide 

 

 

 

 

Our modular, autonomous, turn-key, pulsed Doppler LiDAR systems were conceived to meet the need for remote 
sensing of motion and backscatter in the atmosphere. In this mode of operation, naturally occurring aerosols and 
clouds act as a distributed target and backscatter the transmitted pulses. The receiver detects the Doppler shift 
brought about by the motion of the distributed targets and therefore the line-of-sight component of their velocity. The 
novel optical technology employed and the design approach have led to a new class of eye-safe (Class 1M), high 
performance LiDAR exhibiting exceptional stability which is compatible with a continuous and unattended mode of 
operation. Typical applications include boundary layer wind profiling, plume dispersion, analysis of complex flows, 
cloud studies, cloud base measurements and gust and air quality monitoring. 
 
While every effort has been made to provide accurate and calibrated data products, HALO Photonics does not 
currently guarantee the calibration of the data in absolute terms.   
 
The system has been designed to be rugged and autonomous. Even so, the end user must respect the fact that the 
system is a precision optical instrument that must be treated with great care.

  

 

 
 

The laser emission from the antenna is in the class 1M category. The responsibility for ensuring suitable safety 
procedures and operating modes lies entirely with the end user. HALO Photonics does not accept any 
responsibility for issues relating to the field deployment of the equipment and propagation of the beam in the 
atmosphere. 

Содержание Stream Line XR

Страница 1: ...ntinuous and unattended mode of operation Typical applications include boundary layer wind profiling plume dispersion analysis of complex flows cloud studies cloud base measurements and gust and air q...

Страница 2: ...irus is off Windows time syncing is disabled Daylight saving adjustment is disabled Video resolution is 1280 x 1024 BIOS settings The PC is set to come on with power The PC will come on at 12 20 PM if...

Страница 3: ...to be observed at all times The LiDAR should be sited in such a way as to minimise the amount of heat generated by close surfaces such as tarmac or sunlight falling on it The temperature specification...

Страница 4: ...nsit Case Packing and Unpacking 2 Connections 3 Power requirements and the UPS module 4 Powering up 5 Orientation and levelling 6 Site considerations 7 Routine cleaning 8 Typical thermal characteristi...

Страница 5: ...he appropriate manual handling health and safety guidelines Unpacking reverse the process when packing The transit case consists of 3 separate pieces The LiDAR is packed fully assembled Remove the top...

Страница 6: ...that it be left at this position The small pillars with rubber coned ends act as support legs when the 4 main legs are not attached They prevent the weight of the LiDAR being taken by the cooling agg...

Страница 7: ...coming from the power supply module F 1 10A 5x20mm fuse for the LiDAR circuit F 2 10A 5x20mm fuse for the heaters and the thermo electric cooler TEC circuit if fitted There are 5 small LEDs close to...

Страница 8: ...er un wired smaller connector and the white purging line The arrow on the power connector faces away from the LiDAR when it is the correct orientation The connector is a turn bayonet type Ensure that...

Страница 9: ...e to the UPS and if a power cut happens a warning message will be displayed in the software tab area which will warn that a power cut is detected and the time remaining until the LiDAR shuts down With...

Страница 10: ...e main connector body using a left hand thread Recommended PSU for connection to the 24V DC input power protection module It is recommended that the no load voltage of the PSU be turned up to 25 5V be...

Страница 11: ...internal PC will be fully booted approximately 1 minute later indicated by the flashing green LED on the connector panel cluster If the unit has been allowed to cool down to below 15 C then additional...

Страница 12: ...degrees Note For CSM scan files which are defined in motor encoder counts not degrees the offset bearing will not be used and all movements are with respect to the home position zero counts It will be...

Страница 13: ...the heat generated underneath the LiDAR from direct sunlight a turfed area is ideal or shading if possible Ambient temperatures quoted in the temperature specification of the LiDAR are with respect to...

Страница 14: ...coverage of the clear aperture is not too high The output window has a hard anti reflection coating on the outer surface The coating is reasonably robust but should it be deemed necessary to clean it...

Страница 15: ...n If fuse F 2 is not blown then there may be a fault with the cooling unit in which case the LiDAR should be powered down to avoid the risk of it over heating Passive temperature option The heat excha...

Страница 16: ...passwords needed to access the system are either removed or are sent to Halo prior to sending the system The system is packed in the same way that it was received The Unpacking section can be used for...