imager and measure his temperature with the thermal imager;
If the temperature reading is not greater than 37.2
℃
, the
calibration completes; otherwise the calibration parameter
should be adjusted accordingly.
Step 4: Click “Save” to save the Calibration parameter
4.5. Selection of Color Palette
Two color palette
“
Rainbow
”
and
“
Iron Gray
”
are available for
the users to choose in the
“
Parameter Setting
”
according to their
preference.
Figure 7-Rainbow mode vs Iron Gray mode
4.6. Working modes
The N901 basic version provides only two working mode of
“
single-person temperature measurement mode
”
and
“
large-crowd
temperature measurement mode
”
. While the AI version implements
seven additional modes.The specific working mode can be confirmed
through the "function setting" page in the APP.
14