BASIC SETTINGS
Within the “Cuff Control” menu in [Basic Settings] mode, there are three
options available for configuring your BPSim
™
units:
Heart Rate
This sets the pulses per minute of the needle and Korotkoff sounds
emitted by the cuff just above and below the range between the
Systolic and Diastolic blood pressure values.
Systolic Pressure
Represents the upper pressure threshold at which blood flow within the
artery can first be heard. The 1
st
phase of Korotkoff sounds should begin
right when the pressure gauge falls below this level.
Diastolic Pressure
Represents the lower pressure threshold at which blood flow within the
artery can last be heard. The 4
th
phase of Korotkoff sounds should be
audible immediately above this pressure level, and should completely
disappear (the 5
th
Korotkoff phase) when the pressure gauge falls below
this level.