Instructions for use SURGmatic S11 L - 1.009.1010 SURGmatic S11 C - 1.009.1005
7 Reprocessing steps in accordance with EN ISO 17664 | 7.5 Care products and systems - Servicing
7.4.1 Automated internal and external cleaning and
internal and external disinfection
Note
Prior to cleaning or disinfection in a thermodisinfector, attach the head to an
appropriate shank.
KaVo recommends washer disinfectors according to ISO 15883-1, which are
operated using alkaline cleaning agents.
▶
For programme settings as well as cleansers and disinfectants to be used,
please refer to the Instructions for Use of the washer disinfector.
▶
For the spray tube, also use the adapter for external spray channels.
7.4.2 Automated drying
The drying procedure is normally part of the cleaning programme of the washer
disinfector.
Note
Please comply with the instructions for use of the washer disinfector.
▶
In order to prevent impairment of the KaVo medical device, make sure that
the inside and outside of the device is dry after the end of the cycle.
▶
Immediately after drying, lubricate the KaVo medical device with care
agents from the KaVo care system.
7.5 Care products and systems - Servicing
WARNING
Sharp dental bur or diamond grinder in the medical device.
Risk of injury from sharp and/or pointed dental bur or diamond grinder.
▶
Remove dental bur or diamond grinder.
CAUTION
Premature wear and malfunctions from improper servicing and care.
Reduced product life.
▶
Perform proper care regularly!
Note
KaVo guarantees the proper function of KaVo products only if the care
products listed as accessories are used, since these were tested for proper
use on our products.
7.5.1 Care with KaVo Spray
KaVo recommends servicing the product after each use, i.e. after each auto-
matic cleaning and before each sterilisation.
▶
Remove the dental bur or diamond grinder.
▶
Cover the product with the Cleanpac bag.
17 / 24