Select a suitable procedure (depending on the available autoclave) from
the following sterilisation processes:
▪ Autoclave with three times pre-vacuum:
–
at least 3 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4
°F)
▪ Autoclave using the gravity method:
–
at least 10 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4
°F)
▶
Use according to the manufacturer's Instructions for Use.
7.9 Storage
Prepared products must be stored, protected from germs (as far as pos‐
sible) and dust, in a dry, dark, cool room.
Reprocessing steps in accordance with ISO 17664
59